In Situ Optimization of pH for Parts-Per-Billion Electrochemical Detection of Dissolved Hydrogen Sulfide Using Boron Doped Diamond Flow Electrodes
Authors:
- Maxim B. JosephMJMaxim B. Joseph
- Tania L. ReadTRTania L. Read
- Tim MollartTMTim Mollart
- Mark E. NewtonMNMark E. Newton
- Julie V. MacphersonJMJulie V. Macpherson
- 28 September 2014
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 86 (21) , 10834-10840
Abstract
A novel electrochemical approach to the direct detection of hydrogen sulfide (H2S), in aqueous solutions, covering a wide pH range (acid to alkali), is described. In brief, a dual band electrode device is employed, in a hydrodynamic flow cell, where the upstream electrode is used to controllably generate hydroxide ions (OH–), which flood the downstream detector electrode and provide the correct pH environment for complete conversion of H2S to the electrochemically detectable, sulfide (HS–) ion. All-diamond, coplanar conducting diamond band electrodes, insulated in diamond, were used due to their exceptional stability and robustness when applying extreme potentials, essential attributes for both local OH– generation via the reduction of water, and for in situ cleaning of the electrode, post oxidation of sulfide. Using a galvanostatic approach, it was demonstrated the pH locally could be modified by over five pH units, depending on the initial pH of the mobile phase and the applied current. Electrochemical detection limits of 13.6 ppb sulfide were achieved using flow injection amperometry. This approach which offers local control of the pH of the detector electrode in a solution, which is far from ideal for optimized detection of the analyte of interest, enhances the capabilities of online electrochemical detection systems.Keywords
Funding Information
- University of Warwick
- Element Six
This publication has 46 references indexed in Scilit:
- Capture and Visualization of Hydrogen Sulfide by a Fluorescent ProbeAngewandte Chemie International Edition, 2011
- Fabrication of Versatile Channel Flow Cells for Quantitative Electroanalysis Using PrototypingAnalytical Chemistry, 2010
- Amperometric Oxygen Sensor Based on a Platinum Nanoparticle-Modified Polycrystalline Boron Doped Diamond Disk ElectrodeAnalytical Chemistry, 2008
- Examination of the Spatially Heterogeneous Electroactivity of Boron-Doped Diamond Microarray ElectrodesAnalytical Chemistry, 2006
- Impact of Grain-Dependent Boron Uptake on the Electrochemical and Electrical Properties of Polycrystalline Boron Doped Diamond ElectrodesThe Journal of Physical Chemistry B, 2006
- Application of diamond electrodes to electrochemical processesElectrochimica Acta, 2005
- Electrochemical determination of hydrogen sulfide at carbon nanotube modified electrodesAnalytica Chimica Acta, 2004
- Peer Reviewed: Flow Injection Analysis: From Beaker to Microfluidics.Analytical Chemistry, 2000
- Model-based validation of a DOx sensorControl Engineering Practice, 1996
- Flow injection analysis. principles, applications and trendsAnalytica Chimica Acta, 1980