Abstract
Background: We wished to determine whether newly available flat panel detector (FPD) c-arms were (1) associated with lower radiation dose during ureteroscopy than conventional image intensifier (CII) c-arms, and (2) to compare fluoroscopic image quality between the units. Materials and Methods: We retrospectively reviewed 44 consecutive patients undergoing ureteroscopy at a pediatric hospital, with c-arms assigned by availability in the operating room. We performed dosimetry experiments using the same c-arms on standard phantoms. Results: Patient and case characteristics did not differ significantly between the two groups of patients. The median dose in the FPD group was less than a quarter of the dose in the CII group, 0.48 [0.42, 0.97] mGy versus 2.2 [1.1, 38], p<0.0001. The FPD dose remained at less than 1/3rd of the CII dose accounting for any difference in fluoroscopy time, and remained significant in a multivariate model including fluoroscopy time and patient weight (β=2.4, p=0.007). Phantom studies showed higher image quality for FPDs at all simulated patient sizes, even at lower radiation doses. Conclusions: This is the first report comparing radiation dose from c-arms of image intensifiers and FPDs in adults or children. Use of an FPD during ureteroscopy was associated with a substantially decreased absorbed dose for patients while simultaneously improving image quality.