Abstract
A unique field-effect transistor with an active channel composed of aptamers designed to bind to the Zika virus was fabricated. To immobilize the aptamers in the device channel, we used gold nanoparticles (AuNPs), a 20-40 nm thin film of sputtered gold (Au film), and a molecular sieve material zeolite. The aptamers were functionalized with a thiol end group that enabled them to bind with gold and most other materials. In the case of aptamer/AuNP, when Zika was added to the channel it reduced the channel conductivity from 4.4x102 S/cm to 3.7 S/cm. The gate field effect clearly showed that the aptamer/AuNP channel is p-type with a relatively large transconductance of 300 μS (VDS =2V) compared to the Zika/aptamer/AuNP channel (also p-type) but with lower transconductance of 2 μS (VDS=2V). There was large amount of charge trapping (~1015 cm2) in Zika/aptamer/AuNP channel. In other cases of aptamer/Au film and zeolite channels, there were large changes in the channel conductance when Zika was added. Aptamer/Au and aptamer/zeolite channels were both n-type and their conductivities increased when Zika was added. Combination of AuNP and zeolite biosensing FETs can be used to design complimentary bio-FET sensors. We show that the gate voltage in FET biosensors can be used to increase their sensitivity (24 times in Zika/aptamer/AuNP), selectivity and dynamic range (by 10x or more).Keywords
Funding Information
- NSF (National Science Foundation) EAGER (1931100)
This publication has 42 references indexed in Scilit:
- Enhanced Sensing of Nucleic Acids with Silicon Nanowire Field Effect Transistor BiosensorsNano Letters, 2012
- Do Concentration Cells Store Charge in Water? Comment on Can Water Store Charge?Langmuir, 2009
- One-step RT-PCR for detection of Zika virusJournal of Clinical Virology, 2008
- Dengue Virus Detection Using Whole Blood for Reverse Transcriptase PCR and Virus IsolationJournal of Clinical Microbiology, 2007
- Aptamer-Modified Gold Nanoparticles for Colorimetric Determination of Platelet-Derived Growth Factors and Their ReceptorsAnalytical Chemistry, 2005
- Binding to an RNA Aptamer Changes the Charge Distribution and Conformation of Malachite GreenJournal of the American Chemical Society, 2002
- Miniaturised thin film conductometric biosensors with high dynamic range and high sensitivitySensors and Actuators B: Chemical, 1997
- Conducting polymer-based biosensorsElectrochimica Acta, 1994
- High-Resolution Molecular Discrimination by RNAScience, 1994
- Thin-film conductometric biosensors for glucose and urea determinationBiosensors and Bioelectronics, 1994