Abstract
Digital holographic microscopy can be used for quantitative and dynamic phase imaging of biological cells, which has a very wide application in the fields of biological and medical sciences. To make the measurement system simple, compact, and low-cost, a common-path configuration based on a beam displacer unit is introduced in digital holographic microscopy. The simple optical structure and common-path design reduces the system requirement for the light source coherence, realizes the convenient adjustment of the object and reference beams and achieves an excellent system temporal stability of 0.53 nm. Living mouse osteoblastic cells during mitotic phase are quantitatively measured to demonstrate the capability and applicability of the system.
Funding Information
  • National Natural Science Foundation of China and the China Academy of Engineering Physics NSAF (U1730137)
  • Key Research and Development Program of Shaanxi Province (2017KW-012)

This publication has 27 references indexed in Scilit: