Performance evaluation of a small‐animal PET scanner with 213 mm axis using NEMA NU 4‐2008
Authors:
- 21 October 2024
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 52 (1) , 530-541
Abstract
Background: Long‐axis positron emission tomography (PET) has emerged as one of the recent research directions in PET due to its ability to significantly enhance sensitivity and counting performance for low‐dose imaging, rapid imaging, and whole‐body dynamic imaging.Purpose: The PET system presented in this study is a long‐axis animal PET based on lutetium–yttrium orthosilicate and silicon photomultiplier, designed for whole‐body imaging in rats. It features a diameter of 143 mm and an axial length of 213.3 mm. This study evaluated the performance of this PET system in accordance with the National Electrical Manufacturers Association (NEMA) NU 4‐2008 standards.Methods: The performance evaluation was conducted according to the NEMA NU 4‐2008 standards in terms of spatial resolution, sensitivity, counting rate performance, scatter fraction (SF) and image quality. In addition, a rat imaging study was conducted to assess the imaging capability of this PET system.Results: The average energy resolution of the PET system was 12.87%, the average coincidence timing resolution was 751 ps. The FWHM of spatial resolution reconstructed by filtered back projection and 3D‐OSEM‐PSF algorithm at 5 mm radial offset from the axial center were 1.65 and 0.88 mm. The peak absolute sensitivity measured by a point source at the center of the field of view was evaluated as 6.71% (361–661 keV) and 10.31% (250–750 keV). For the mouse‐like phantom, the SF was 11.0% and the peak noise equivalent counting rate (NECR) was 1193 kcps at 94.2 MBq (2.54 mCi). For the rat‐like phantom, the SF was 26.8% and the NECR was 682.5 kcps at 78.6 MBq (2.12 mCi).Conclusions: The performance measurement results demonstrate that this PET system exhibits high sensitivity and count rate performance, making it potential for high‐quality whole‐body dynamic imaging of rats.Keywords
Funding Information
- National Key Research and Development Program of China (2022YFF0707900)
- National Natural Science Foundation of China (12275291, 12175265, 12075267)
- Natural Science Foundation of Shandong Province (ZR2021MA106)
This publication has 38 references indexed in Scilit:
- NEMA NU 2–2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CTEuropean Journal of Nuclear Medicine and Molecular Imaging, 2022
- Advances in Preclinical PET InstrumentationPET Clinics, 2020
- Applied Systems Biology—embracing molecular imaging for systemic medicineEuropean Journal of Nuclear Medicine and Molecular Imaging, 2020
- The MINDView brain PET detector, feasibility study based on SiPM arraysNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- DigiPET: sub-millimeter spatial resolution small-animal PET imaging using thin monolithic scintillatorsPhysics in Medicine & Biology, 2014
- Study of PET scanner designs using clinical metrics to optimize the scanner axial FOV and crystal thicknessPhysics in Medicine & Biology, 2013
- NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomographPhysics in Medicine & Biology, 2013
- Design and performance evaluation of a compact, large-area PET detector module based on silicon photomultipliersNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2012
- Effective count rates for PET scanners with reduced and extended axial field of viewPhysics in Medicine & Biology, 2011
- Performance evaluation of an Inveon PET preclinical scannerPhysics in Medicine & Biology, 2009