Welder identity, weld date, and the risk of outlet strut fracture in Björk-Shiley convexo-concave valves: the Dutch cohort study.
Authors:
- A. AlgraAAA. Algra
- Y. van der GraafYGY. van der Graaf
Open Access
Abstract
OBJECTIVE: To establish whether there is an association between subsequent fracture of the outlet strut and welder identity and the weld date of Björk-Shiley convexo-concave (BScc) valves. DESIGN: Cohort study. PATIENTS: All Dutch BScc valve recipients (n = 2266). MEAN OUTCOME MEASURES: Documented outlet strut fracture during follow up. RESULTS: Weld dates were known for 97.0% of all BScc valves (n = 2534) implanted in Dutch patients (n = 2266) and welder identity was known for 52.2%. During a mean follow up of 9.4 years, 46 fractures were documented. For 60 degrees valves welded from 1981 to 1984 the fracture rate (0.22 per 100 person-years (95% CI 0.13 to 0.34)) was higher than that for valves welded before 1981 (0.04 (95% CI 0.01-0.10)). When all fracture related risk factors were taken into account, fracture rates per welder did not show any statistically significant differences. CONCLUSIONS: Welder identity does not contribute to the risk of strut fracture. Other factors in the production of BScc valves may explain the increased risk for valves welded from 1981 to 1984.Keywords
This publication has 7 references indexed in Scilit:
- Single-leg strut fractures in explanted Björk-Shiley valvesThe Lancet, 1994
- Risk of strut fracture of Björk-Shiley valvesThe Lancet, 1992
- Lethal Blockage of a Bjørk-Shiley Artificial Heart Valve Caused by Strut Fracture - the Metallurgical AspectThe Thoracic and Cardiovascular Surgeon, 1989
- MECHANICAL FAILURE OF THE BJORK-SHILEY VALVE - UPDATED FOLLOW-UP AND CONSIDERATIONS ON PROPHYLACTIC REREPLACEMENT1989
- Outlet strut fracture of the Björk-Shiley 60 ° Convexo-Concave valve: Current information and recommendations for patient careJACC, 1988
- The Björk-Shiley 70° Convexo-Concave Prosthesis Strut Fracture ProblemThe Thoracic and Cardiovascular Surgeon, 1987