Chemical analysis of impurities in zirconia and aluminium nitride ceramics by inductively coupled plasma atomic emission spectrometry
Authors:
- Charo Martínez-LebrusantCMCharo Martínez-Lebrusant
- Flora BarbaFBFlora Barba
Open Access
- 1 January 1990
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 115 (10) , 1335-1338
Abstract
Inductively coupled plasma atomic emission spectrometry (ICP-AES) has been used for the determination of the most significant impurities (Al, Ca, Fe, Mg, P, Si and Ti) in zirconia and aluminium nitride ceramics according to the following procedure: (i) development of two decomposition methods for each material; (ii) study of spectral interferences, matrix and interelemental effects; and (iii) determination of the detection limits. Good precision and accuracy were attained. The procedures developed in this work can be applied to zirconia partially stabilised by the addition of CaO, MgO or Y2O3, tetragonal zirconia polycrystals with the addition of Y2O3, pure aluminium nitride or aluminium nitride with additions of alkaline earth oxides, Y2O3 or CeO2.Keywords
This publication has 5 references indexed in Scilit:
- ZrO2 Powders for Advanced and Engineered Ceramics1985
- Development of a high-resolution inductively-coupled argon plasma apparatus for derivative spectrometry and its application to the determination of hafnium in high-purity zirconium oxideTalanta, 1982
- Conversion of “Tables of Spectral-Line Intensities” for NBS copper arc into table for inductively coupled argon plasmasSpectrochimica Acta Part B: Atomic Spectroscopy, 1981
- The determination of small amounts of aluminium and silicon in stabilized zirconia ceramicsAnalytica Chimica Acta, 1978
- New method for decomposition and comprehensive analysis of silicates by atomic absorption spectrometryAnalytical Chemistry, 1968