Modified Technique for the Implantation of Berlin Heart Excor Ventricular Assist Device in Children
- 2 July 2012
- journal article
- research article
- Published by SAGE Publications in World Journal for Pediatric and Congenital Heart Surgery
- Vol. 3 (3) , 373-377
Abstract
The Berlin Heart Excor (BHE) assist device has become our standard mechanical support device for long-term support in children with heart failure. We report two useful surgical modifications for the implantation of the BHE in the pediatric population, improving ease of implantation as well as subsequent surgical procedures at the time of explantation or transplantation. The first modification entails the use of a polytetrafluoroethylene graft for cannulation via the innominate artery for arterial perfusion during cardiopulmonary bypass. This graft is preserved and reused for institution of bypass at the time of transplantation or explantation of the device after recovery. The second modification consists of an extension of the BHE arterial cannula using a length of knitted polyester graft. This allows for improved positioning of the arterial cannula in the ascending aorta or main pulmonary artery, facilitates vascular anastomosis, and improves hemostasis.Keywords
This publication has 5 references indexed in Scilit:
- A technique for implanting outflow cannulas for Berlin Heart EXCOR ventricular assist device in small pediatric patientsThe Journal of Thoracic and Cardiovascular Surgery, 2011
- Single-center experience with treatment of cardiogenic shock in children by pediatric ventricular assist devicesThe Journal of Thoracic and Cardiovascular Surgery, 2011
- Bridging children of all sizes to cardiac transplantation: The initial multicenter North American experience with the Berlin Heart EXCOR ventricular assist deviceThe Journal of Heart and Lung Transplantation, 2011
- Ventricular assist device implantation in neonates: Adjustment of the BerlinHeart EXCOR arterial cannula with bovine pericardiumThe Journal of Thoracic and Cardiovascular Surgery, 2010
- Surgical strategy to prevent cardiac injury during reoperation in infantsJournal of Cardiothoracic Surgery, 2008