Abstract
Carbon microfibers (MFs) coated with conducting polymers may provide a solution for long-term recording of activity from individual or small groups of neurons. Attaching cell adhesion molecules to the electro-sensitive surface might further improve electrode–neuron contact, thus enhancing signal stability and fidelity. We fabricated biofunctionalized microelectrodes consisting of 7-μm diameter carbon MFs coated with poly(3,4-ethylenedioxythiophene) doped with poly[(4-styrenesulfonic acid)-co-(maleic acid)] (PEDOT:PSS-co-MA), and linked N-Cadherin to the polymer surface. These electrodes were tested for recording artificially generated electric potentials, as well as multiunit activity (MUA), sharp wave-ripple complexes (SWRs), and field excitatory postsynaptic potentials (fEPSPs) in rat hippocampal slices. The effects of electrode length and functionalization were compared. PEDOT:PSS-co-MA coating improved electric current detection and reduced the electrical noise but had no significant effect on the amp...
Funding Information
  • Instituto de Salud Carlos III
  • Ministerio de Economía y Competitividad (PI12/02835)
  • Eropean Regional Development Fund (PI12/02835)

This publication has 47 references indexed in Scilit: