Assessing the Adequacy of Peripherally Inserted Central Catheters for Power Injection of Intravenous Contrast Agents for CT
Authors:
- Eric E. WilliamsonEWEric E. Williamson
- J. Mark McKinneyJMJ. Mark McKinney
- 1 November 2001
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 25 (6) , 932-937
Abstract
The purpose of this work was to determine the tolerance of silicone peripherally inserted central catheters (PICCs) of different sizes and lengths to power injection of contrast materials at flow rates suitable for CT studies. Fifty silicone PICCs in three single-lumen sizes (3 to 5F) and two double-lumen sizes (6 and 7F) were cut to two lengths (35 and 45 cm), and a uniform volume of 74% ioversol was injected into each at increasing rates of flow by a power injector. The flow rate, volume, and peak pressure were recorded for each injection. The respective tolerated flows for the 35 and 45 cm PICCs were 0.65 ml/s at 125 psi and 0.56 ml/s at 125 psi for the 3F catheters, 1.58 ml/s at 150 psi and 1.04 ml/s at 150 psi for the 4F catheters, 4.20 ml/s at 200 psi and 3.02 ml/s at 170 psi for the 5F catheters, 1.50 ml/s at 145 psi and 0.88 ml/s at 150 psi for the 6F catheters, and 9.52 ml/s at 350 psi and 8.78 ml/s at 330 psi for the 7F catheters. The 3F catheters were unsuitable for power injection for CT studies because they could not accommodate adequate flow rates. The 4F single-lumen and 6F double-lumen catheters withstood flow rates that were marginally adequate for CT studies. The 5F single-lumen and 7F double-lumen PICCs tolerated peak flows and pressures well within the range necessary to allow power injection of contrast materials for CT studies. For each size of PICC, the 35 cm length withstood higher flow rates than the 45 cm length before failure.Keywords
This publication has 10 references indexed in Scilit:
- Power Injection of Peripherally Inserted Central CathetersJournal of Vascular and Interventional Radiology, 1997
- In-line pressures generated in small-bore central venous catheters during power injection of CT contrast media.Radiology, 1997
- Peripherally Inserted Central Catheters in General MedicineMayo Clinic Proceedings, 1997
- Power injection of intravenous contrast material through central venous catheters for CT: in vitro evaluation.Radiology, 1996
- Safe use of power injectors with central and peripheral venous access devices for pediatric CTPediatric Radiology, 1996
- Peripherally inserted central catheters in children.Radiology, 1995
- Interventional Radiologic Placement of Peripherally Inserted Central CathetersJournal of Vascular and Interventional Radiology, 1993
- Placement and management of long-term central venous access catheters and ports.American Journal of Roentgenology, 1993
- Safety Considerations in the Power Injection of Contrast Media Via Central Venous Catheters during Computed Tomographic ExaminationsInvestigative Radiology, 1992
- Safe use of an intravenous power injector for CT: experience and protocol.Radiology, 1990