Abstract
Objective: To enable accurate γ-aminobutyric acid (GABA) quantification from low signal averages in MEscher-GArwood Point-Resolved SpectroScopy (MEGA-PRESS), thereby reducing acquisition time and improving clinical applicability. Methods: A spectroscopy-oriented one-dimensional multichannel Vision Transformer was proposed for GABA+ signal reconstruction from low number of signal averages (low-NSA, 10 transients) MEGA-PRESS magnetic resonance spectra (MRS). The model removes the classification token while preserving all spectral tokens and incorporates a dedicated spectral decoder. A dual-source semisynthetic (DSS) dataset generation strategy was designed, where weighted combinations of high-NSA spectra (0.00-5.00 ppm) and their corresponding fitted GABA+ signals (2.79-3.20 ppm) were used to construct input and labels in the training data. Results: Using high-NSA (NSA160) quantification results as the reference, the proposed method reduced the mean absolute percentage error (MAPE) of the quantification results from 13.32 ± 3.46% (with conventional NSA10) to 3.78 ± 1.46%, with a mean reduction of 9.53% (95% CI: 7.09%–11.97%, p < 0.05). Conclusion: The proposed Transformer-based framework enables accurate reconstruction of GABA+ signals from low-NSA MEGA-PRESS data, preserving both quantitative accuracy and spectral shape. Significance: This approach substantially reduces acquisition times and facilitates faster, more practical in vivo GABA quantification in clinical and research settings.