Response of fourier-transform spectrometers to absorption and emission in a homogeneous single-layered beam splitter.
Authors:
- 1 October 2001
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 40 (28) , 5078-5087
Abstract
Thermal self-emission contributes to interferograms measured with Fourier-transform infrared spectrometers. If the beam-splitter is almost transparent, the complex spectral amplitude that is due to the detector port emission is opposite that of the input port, whereas the amplitude that is due to the beam-splitter emission is in quadrature. The situation of an absorbing beam splitter is examined here. The volume beam-splitter emission is modeled by a superposition of dipole sources spread in an absorbing film. Angular polarization correlations are taken into account. It is found that the phase relations between the complex spectral amplitudes are affected. Numerical data are given for experimental conditions adapted to those of the airborne limb sounder MIPAS-FT.Keywords
This publication has 5 references indexed in Scilit:
- Thermal emission from dielectric beam splitters in Michelson interferometers: a schematic analysisApplied Optics, 2000
- Effect of beam-splitter emission in Fourier-transform emission spectroscopy.Applied Optics, 1999
- Beam splitter layer emission in fourier-transform infrared interferometers.Applied Optics, 1998
- Phase corrections for the emission sounder MIPAS-FTApplied Optics, 1993
- Radiometric calibration of IR Fourier transform spectrometers: solution to a problem with the High-Resolution Interferometer SounderApplied Optics, 1988