Validation of a Novel 3‐Dimensional Sonographic Method for Assessing Gastric Accommodation in Healthy Adults
- 20 May 2016
- journal article
- research article
- Published by Wiley in Journal of Ultrasound in Medicine
- Vol. 35 (7) , 1411-1418
Abstract
Objectives—A novel automated 3-dimensional (3D) sonographic method has been developed for measuring gastric volumes. This study aimed to validate and assess the reliability of this novel 3D sonographic method compared to the reference standard in 3D gastric sonography: freehand magneto-based 3D sonography. Methods—A prospective study with 8 balloons (in vitro) and 16 stomachs of healthy volunteers (in vivo) was performed. After a 500-mL liquid meal, 1 preprandial and 3 postprandial volume scans of the stomachs were performed by the novel 3D sonographic method and the current reference-standard 3D sonographic method. Results—The in vitro study showed a mean volume difference between the novel method and the true balloon volume of –1.3 mL; limits of agreement (LoA) were small (–39.3 to12.3 mL), with an intraclass correlation coefficient (ICC) of 0.998. The in vivo study showed a mean gastric volume of 321 mL between the novel method and the freehand magneto-based method, with a mean volume difference of –4.4 mL; LoA were –40.1 to 31.2 mL, and the ICC was 0.991. The intraobserver and interobserver variability rates were low, at 0.8 mL (LoA, –24.0 to 25.6 mL), with an ICC of 0.995, and 0.5 mL (LoA, of –26.8 to 27.8 mL), with an ICC of 0.999, respectively. Conclusions—The novel 3D sonographic method with automated acquisition showed good agreement with the current reference-standard gastric 3D sonographic method, with low intraobserver and interobserver variability. This novel 3D sonographic method is a valid and reliable technique for determining gastric accommodation.Keywords
This publication has 23 references indexed in Scilit:
- Gastric volume changes in response to a meal: Validation of magnetic resonance imaging versus the barostatJournal of Magnetic Resonance Imaging, 2011
- Defining Left Ventricular Apex-to-Base Twist Mechanics Computed From High-Resolution 3D EchocardiographyJACC: Cardiovascular Imaging, 2010
- Validation of 3D Ultrasound Vessel Wall Volume: An Imaging Phenotype of Carotid AtherosclerosisUltrasound in Medicine & Biology, 2007
- Gastric accommodation and motility are influenced by the barostat device: assessment with magnetic resonance imagingAmerican Journal of Physiology-Gastrointestinal and Liver Physiology, 2007
- Fundal dysaccommodation in functional dyspepsia: head-to-head comparison between the barostat and three-dimensional ultrasonographic techniqueGut, 2006
- Assessment of gastric accommodation: overview and evaluation of current methodsNeurogastroenterology & Motility, 2004
- Non-invasive measurement of gastric accommodation in humansGut, 2002
- Three-Dimensional Ultrasonography of the Gastric Antrum in Patients with Functional DyspepsiaScandinavian Journal of Gastroenterology, 1996
- In vitro evaluation of three-dimensional ultrasonography in volume estimation of abdominal organsUltrasound in Medicine & Biology, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986