Validation of a new panel of automated chemiluminescence assays for von Willebrand factor antigen and activity in the screening for von Willebrand disease
- 3 April 2013
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 35 (5) , 555-565
Abstract
Summary: Introduction: The diagnosis of von Willebrand disease (VWD) largely depends on the results of von Willebrand factor (VWF) antigen and activity. Recently, a new automated VWF:RCo assay on Acustar was developed. This assay panel for VWD also contains a new antigen (VWF:Ag) test. In this study, both chemiluminescence tests (HemosIL VWF:Ag and VWF:RCo) were evaluated.Materials and Methods: Imprecision, limit of detection (LOD), and linearity were evaluated. Method comparison (with VWF:Ag latex assay and VWF:RCo by aggregometry) was performed and diagnostic performance of the new test panel was examined.Results: The imprecision was 7%, and the LOD was 0.2 IU/dL for both assays. Dilution series showed a large linearity for both HemosIL VWF:Ag (0–300 IU/dL) and VWF:RCo (0–200 IU/dL) and method comparison studies revealed good agreement with the currently used VWD panel. The new panel showed adequate diagnostic performance: diagnostic sensitivity was 100% and diagnostic specificity 82% compared with the VWF:Ag latex assay and VWF:RCo by aggregometry. In addition, the new HemosIL Acustar VWF:Ag and HemosIL Acustar VWF:RCo are more sensitive for VWD than the currently used assays.Conclusions: This new VWD test panel has adequate laboratory characteristics and allows fully automated and simultaneous analysis of the VWF:Ag and VWF:RCo.Keywords
This publication has 35 references indexed in Scilit:
- Assessment ofvon Willebrand Factor Propeptide Improves the Diagnosis ofvon Willebrand DiseaseSeminars in Thrombosis and Hemostasis, 2011
- von Willebrand Disease: Local Diagnosis and Management of a Globally Distributed Bleeding DisorderSeminars in Thrombosis and Hemostasis, 2011
- Validation of improved performance characteristics for the automated von Willebrand factor ristocetin cofactor activity assayJournal of Thrombosis and Haemostasis, 2010
- Genetic testing for von Willebrand disease: the case againstJournal of Thrombosis and Haemostasis, 2010
- Genetic testing for von Willebrand disease: the case forJournal of Thrombosis and Haemostasis, 2010
- Identification, Pathogenesis, and Treatment of Factor InhibitorsSeminars in Thrombosis and Hemostasis, 2009
- Clinical and laboratory diagnosis of von Willebrand disease: A synopsis of the 2008 NHLBI/NIH guidelinesAmerican Journal of Hematology, 2009
- A highly-sensitive plasma von Willebrand factor ristocetin cofactor (VWF:RCo) activity assay by flow cytometryJournal of Thrombosis and Haemostasis, 2008
- Is the activated partial thromboplastin time suitable to screen for von Willebrand factor deficiencies?Blood Coagulation & Fibrinolysis, 2007
- Provisional criteria for the diagnosis of VWD type 1: on behalf of the ISTH SSC Subcommittee on von Willebrand FactorJournal of Thrombosis and Haemostasis, 2005