Respiratory Sensing Lead Malfunction in Upper Airway Stimulation: A Single Institution Report
Open Access
- 21 September 2023
- journal article
- research article
- Published by Wiley in The Laryngoscope
- Vol. 134 (3) , 1479-1484
Abstract
Objectives: Upper airway stimulation (UAS) is an effective treatment for patients with obstructive sleep apnea. The device consists of a stimulating electrode, an internal pulse generator, and a respiratory sensing lead. The purpose of this study is to characterize the incidence of sensing lead malfunction necessitating revision surgery in a high‐volume center in conjunction with a review of the FDA Manufacturer and User Facility Device Experience (MAUDE) database reports on adverse events associated with the sensing lead component.Methods: Patients age ≥18 implanted with the device were reviewed between July 2017 and June 2022. Those determined to have sensing lead malfunction were analyzed. The FDA MAUDE database was reviewed for reports associated with “Inspire Model 4323 Respiratory” and “Inspire Model 4340 Respiratory” between January 2014 and September 2022.Results: One hundred ninety patients underwent UAS at our institution during the study period and four (2.1%) patients were found to have sensing lead malfunction. Analysis of the MAUDE database revealed 122 reports of adverse events associated with the sensing lead component, with 72% (88/122) of these requiring revision surgery. The most frequently reported adverse event was sensing lead tip separation from lead body, noted in 46% of these reports (56/122).Conclusion: The functionality of the sensing lead component is vital to the integrity of the UAS device and is measurable through waveform analysis. To provide optimal care after device implantation, it is essential to have a comprehensive understanding of the scenarios in which this component fails, the incidence of such events, and appropriate management.Level of Evidence: 4 Laryngoscope, 134:1479–1484, 2024Keywords
This publication has 13 references indexed in Scilit:
- Space invader: pleural penetration of a hypoglossal nerve stimulator sensor leadJournal of Clinical Sleep Medicine, 2021
- 819 A RARE CASE OF RESPIRATORY SENSE LEAD DYSFUNCTION IN A PATIENT WITH A HYPOGLOSSAL NERVE STIMULATORSLEEP, 2021
- Hypoglossal Nerve Stimulation in Veterans With Comorbid Insomnia and Sleep ApneaOtolaryngology–Head and Neck Surgery, 2021
- Hypoglossal Nerve Stimulation Therapy for the Treatment of Obstructive Sleep ApneaNeurotherapeutics, 2021
- Adverse events associated with the Inspire implantable hypoglossal nerve stimulator: A MAUDE database reviewAmerican Journal of Otolaryngology, 2020
- Adverse Events in Hypoglossal Nerve Stimulator Implantation: 5‐Year Analysis of the FDA MAUDE DatabaseOtolaryngology–Head and Neck Surgery, 2020
- Hypoglossal nerve stimulator implantation via a 2-incision techniqueOperative Techniques in Otolaryngology-Head and Neck Surgery, 2020
- Changes in breath cycle sensing affect outcomes in upper airway stimulation in sleep apneaLaryngoscope Investigative Otolaryngology, 2020
- Upper Airway Stimulation for Obstructive Sleep Apnea: 5‐Year OutcomesOtolaryngology–Head and Neck Surgery, 2018
- Hypoglossal nerve stimulation for obstructive sleep apnea: A review of the literatureWorld Journal of Otorhinolaryngology - Head and Neck Surgery, 2016