Benchtop testing and comparisons among three types of through-the-scope endoscopic clipping devices
Authors:
- 5 January 2013
- journal article
- research article
- Published by Springer Nature in Surgical Endoscopy
- Vol. 27 (5) , 1521-1529
Abstract
Through-the-scope (TTS) endoscopic clipping devices are widely used. No benchtop testing or direct comparisons of these endoclips have been performed to show their rotational ability and inherent mechanical strengths during closure and after deployment. This study aimed to provide benchtop data that can be used to guide clinical applications and to promote future device research and development. Benchtop testing and comparisons were performed for three groups of TTS clips: QuickClip2 long, resolution, and instinct clips. The main outcome measurements were device-in-endoscope retroflection angles (DIERA), opening strength, “snapping” force of acute clip closure, and neoprene pulling strength. The achievable gastroscope DIERA was 10° for QuickClip2, 3° for the resolution clip, and 10° for the instinct clip. The QuickClip and the Instinct clip rotated almost equally well under all endoscope configurations, including endoscopic retrograde cholangiopancreatography (ERCP). With or without a sheath, the resolution clip lacked the ability to rotate. During clip opening force testing (the amount of force required to force open the jaws of a deployed clip by 3.2 mm; 3.2 mm was chosen due to the standard dimension of the gauge used for the measurement), the Instinct clips were the strongest. For the Instinct clips, an opening force of 404 ± 124 g was needed to open the closed clip, and an additional 386 ± 133 g was required to open the clip jaws to 3.2 mm. In terms of snapping force during acute closure and neoprene pulling strength, the instinct and resolution clips performed almost equally. The limitations of the study were the benchtop testing and the finite sample size for closing and pulling strength comparisons. The QuickClip2 and the Instinct clip rotate equally well under different endoscope configurations. The resolution clips lack rotational ability. The instinct clips are stronger mechanically than the other two TTS clips. Stronger clips are perhaps associated with higher therapeutic efficacy and retention rates.Keywords
This publication has 15 references indexed in Scilit:
- Prospective, randomized comparison of 3 different hemoclips for the treatment of acute upper GI hemorrhage in an established experimental settingGastrointestinal Endoscopy, 2012
- Endoscopic full-thickness GI wall resection: current statusGastrointestinal Endoscopy, 2012
- Endoscopic hemostatic devicesGastrointestinal Endoscopy, 2009
- Endoscopic mechanical hemostasis of GI arterial bleeding (with videos)Gastrointestinal Endoscopy, 2007
- Comparative study of endoscopic clips: duration of attachment at the site of clip applicationGastrointestinal Endoscopy, 2007
- Randomized controlled study of 3 different types of hemoclips for hemostasis of bleeding canine acute gastric ulcersGastrointestinal Endoscopy, 2006
- “Hemodynamic Efficacy” of two Endoscopic Clip Devices Used in the Treatment of Bleeding Vessels, Tested in an Experimental Setting Using the Compact Erlangen Active Simulator for Interventional Endoscopy (compactEASIE) Training ModelEndoscopy, 2006
- A randomised controlled comparison of injection, thermal, and mechanical endoscopic methods of haemostasis on mesenteric vesselsGut, 1998
- Titanium Aneurysm Clips: Mechanical Characteristics and Clinical TrialNeurologia medico-chirurgica, 1998
- A Comparative Study of Intracranial Aneurysm Clips: Closing and Opening Forces and Physical EnduranceNeurosurgery, 1997