Design and performance of an evolved gas analysis ion attachment mass spectrometer
Authors:
- 4 August 2010
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 24 (17) , 2625-2630
Abstract
We designed a simple evolved gas analysis (EGA) system to act as a sampler between solid samples at atmospheric pressure and the high vacuum inside a mass spectrometer. The newly designed stainless steel system is simple, small and rugged and fulfills all the basic requirements for EGA. The temperature is programmable with 60°C/min as the maximum heating rate and the temperature range is up to 600°C. With this system coupled with lithium ion attachment mass spectrometry (IAMS), it is possible to study the temperature‐programmed decomposition of a number of solid materials by detecting any chemical species on a real‐time basis. For illustrative purposes, EGA‐IAMS experiments of polyethylene polymers have been conducted. Copyright © 2010 John Wiley & Sons, Ltd.Keywords
This publication has 22 references indexed in Scilit:
- Ion Attachment Mass Spectrometry Combined with Infrared Image Furnace for Thermal Analysis: Evolved Gas Analysis StudiesAnalytical Chemistry, 2009
- Application of Ion Attachment Mass Spectrometry to Evolved Gas Analysis for in Situ Monitoring of Porous Ceramic ProcessingAnalytical Chemistry, 2006
- Reexamination of the Pyrolysis of Polyethylene: Data Needs, Free-Radical Mechanistic Considerations, and Thermochemical Kinetic Simulation of Initial Product-Forming PathwaysMacromolecules, 2003
- Lithium ion attachment mass spectrometry: Instrumentation and featuresReview of Scientific Instruments, 2001
- Alkali-metal ion/molecule association reactions and their applications to mass spectrometryMass Spectrometry Reviews, 2000
- Chemical ionization mass spectrometry with lithium ion attachment to the moleculeAnalytical Chemistry, 1989
- Evolved gas analysis by mass spectrometry: A reviewThermochimica Acta, 1984
- Investigation of short-chain branches in polyethylene by pyrolysis - GCMSMacromolecules, 1983
- A temperature programmed fractionation inlet system for mass spectrometersAnalytical Chemistry, 1982
- An automated system for simultaneous thermal analysis and mass spectrometry. Part 1Thermochimica Acta, 1982