Influence of temperature on the frequency response of fine-wire thermocouples over the range (300K-800K) in airflows
Authors:
- 1 September 1984
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 17 (9) , 808-812
Abstract
Theoretical and experimental investigation of the frequency response of fine-wire thermocouples is presented for the case of heated airflows (300K-1-1. The influence of temperature on the thermocouple transfer function has been studied through the variation of physical properties of surrounding gas and thermocouple material. Dynamic calibration has been carried out by using an 'external heating technique'. This study provides evidence that, over the velocity and temperature ranges considered, the transfer function depends on thermocouple geometry (rw and rb/rw) as well as on characteristics of the flow (velocity and temperature).Keywords
This publication has 3 references indexed in Scilit:
- Frequency response of fine wire thermocoupleJournal of Physics E: Scientific Instruments, 1982
- A New On-Line Pulsing Technique for Response Measurements of Thermocouple WiresCombustion Science and Technology, 1981
- Heat transfer from spheres to flowing mediaPhysica, 1946