Intra‐Rater Variability and Inter‐Rater Reliability in Measuring Tear Osmolarity in Dogs Using the ScoutPro System
- 24 July 2025
- journal article
- Published by Wiley in Veterinary Ophthalmology
- Vol. 28 (6) , 970-976
Abstract
Objective: To evaluate the applicability of the ScoutPro (Trukera Medical, Southlake, Texas, USA) osmolarity system for tear film osmolarity measurement in healthy dogs. Animals Studied: Seventeen dogs (34 eyes) of dolichocephalic and mesocephalic breeds were tested. Procedures: Tear osmolarity was tested with the ScoutPro osmolarity system. Two examiners performed three subsequent measurements of each eye. Intra‐rater variability and inter‐rater reliability were calculated. Results: The technique was straightforward; instrument handling was simple. Mean tear osmolarity measured by examiner one was 323.51 ± 9.89 mOsm/L and by examiner two 326.74 ± 9.02 mOsm/L. Mean difference between examiners was 3.2 ± 11.2 mOsm/L. Paired two‐tailed t ‐test was p = 0.1. The intra‐rater variability was acceptable with a Cronbach's Alpha of 0.75 for rater one and 0.80 for rater two. There were no significant differences in the three measurements of the individual examiners. There were no significant differences in the three measurements of the individual examiners (asymptotic significance rater one was 0.56, asymptotic significance rater two was 0.11). Conclusions: Measurements with the ScoutPro osmolarity system in dogs are easy to perform. Variability and inter‐rater reliability seem to be acceptable in the small group of dogs tested. Further studies regarding the reliability of the test results and the evaluation of the tear osmolarity in ocular surface disease are warranted.Keywords
This publication has 19 references indexed in Scilit:
- Diurnal variations of tear film osmolarity on the ocular surfaceClinical and Experimental Optometry, 2022
- Comprehensive Clinical, Diagnostic, and Advanced Imaging Characterization of the Ocular Surface in Spontaneous Aqueous Deficient Dry Eye Disease in DogsCornea, 2019
- Validity and Reliability of a Novel Handheld Osmolarity System for Measurement of a National Institute of Standards Traceable SolutionCornea, 2018
- TFOS DEWS II pathophysiology reportThe Ocular Surface, 2017
- TFOS DEWS II Definition and Classification ReportThe Ocular Surface, 2017
- TFOS DEWS II Tear Film ReportThe Ocular Surface, 2017
- TFOS DEWS II Diagnostic Methodology reportThe Ocular Surface, 2017
- Assessment of tear film osmolarity using the TearLab™osmometer in normal dogs and dogs with keratoconjunctivitis siccaVeterinary Ophthalmology, 2016
- Tear Osmolarity in the Diagnosis and Management of Dry Eye DiseaseAmerican Journal of Ophthalmology, 2011
- Abstracts: Annual Meeting of the European College of Veterinary Ophthalmologists/European Society of Veterinary Ophthalmologists, Copenhagen, Denmark, June 3–7, 2009Veterinary Ophthalmology, 2009