Chemical speciation of nanoparticles surrounding metal-on-metal hips
- 13 June 2012
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Chemical Communications
- Vol. 48 (67) , 8335-8337
Abstract
Spectromicroscopy of tissue surrounding failed CoCr metal-on-metal hip replacements detected corroded nanoscale debris in periprosthetic tissue in two chemical states, with concomitant mitochondrial damage. The majority of debris contained Cr3+, with trace amounts of oxidised cobalt. A minority phase containing a core of metallic chromium and cobalt was also observed.Keywords
This publication has 16 references indexed in Scilit:
- Failure of the Durom Metasul® Acetabular ComponentClinical Orthopaedics and Related Research, 2010
- Nanoparticles can cause DNA damage across a cellular barrierNature Nanotechnology, 2009
- The Epidemiology of Bearing Surface Usage in Total Hip Arthroplasty in the United StatesJournal of Bone and Joint Surgery, 2009
- The painful metal-on-metal hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2009
- MRI of early symptomatic metal-on-metal total hip arthroplasty: a retrospective review of radiological findings in 20 hipsClinical Radiology, 2008
- Neo-capsule tissue reactions in metal-on-metal hip arthroplastyActa Orthopaedica, 2007
- Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulationJournal of Materials Science: Materials in Medicine, 2006
- Metal-on-Metal Bearings and Hypersensitivity in Patients with Artificial Hip JointsJournal of Bone and Joint Surgery, 2005
- Comparison of in vitro with in vivo characteristics of wear particles from metal–metal hip implantsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2004
- Contamination by metallic elements released from joint prosthesesMedical Engineering & Physics, 2003