Early Experience with a Newly Developed Electrospun Polycarbonate-urethane Vascular Graft for Hemodialysis Access
- 21 January 2013
- journal article
- research article
- Published by SAGE Publications in The Journal of Vascular Access
- Vol. 14 (3) , 252-256
Abstract
Purpose: In this pilot study, we tested a newly developed electrospun multilayered, self-sealing graft, AVflo™, specifically designed for early hemodialysis access. Methods: Ten eligible consenting patients had a polycarbonate-urethane graft (AVflo™) implanted and were followed up prospectively for at least six months. Performance measures included graft patency, complications and time to first cannulation. Mean age of the patients was 66.7 ± 10 years. Chronic glomerulonephritis was the most common cause of renal failure. A total of 70% of the patients had a history of previous vascular access and 40% history of minimally invasive radiologic procedures for patency maintenance. In 40% of the cases the need for AV graft implantation was because of recurrent infections from permanent catheter for dialysis. Seven grafts were placed in the upper arm and three in the thigh. Mean follow up was 230 ± 75 days. Results: There were no systemic or local reactions to the graft and we did not report any graft infections. Two grafts thrombosed because of severe bleeding post-cannulation due to an incorrect needle puncture. Both grafts were successfully thrombectomized. Primary and secondary patency rates at six months were 60% and 78%, respectively. These patency rates were comparable to those reported for other polyether-urethane and ePTFE grafts. Median time to first cannulation was seven days (3-21) and all puncture sites sealed in less than five minutes. Conclusions: This newly developed electrospun polycarbonate-urethane graft is safe in humans, permits early access obviating the need for venous catheters, and has equivalent patency as other prosthetic grafts.Keywords
This publication has 25 references indexed in Scilit:
- The State of Chronic Kidney Disease, ESRD, and Morbidity and Mortality in the First Year of DialysisClinical Journal of the American Society of Nephrology, 2009
- Population Requirements for Vascular Access SurgeryEuropean Journal of Vascular and Endovascular Surgery, 2006
- Superiority of Autogenous Arteriovenous Hemodialysis Access: Maintenance of Function with Fewer Secondary InterventionsAnnals of Vascular Surgery, 2004
- Upper Arm Polytetrafluoroethylene Grafts for Dialysis Access. Analysis of Two Different Graft Sizes: 6 mm and 6–8 mmVascular and Endovascular Surgery, 2003
- Clinical and Doppler ultrasonography data of a polyurethane vascular access graft for haemodialysis: a prospective studyNephrology Dialysis Transplantation, 2003
- Improving the patency of vascular bypass grafts: The role of suture materials and surgical techniques on reducing anastomotic compliance mismatchEuropean Journal of Vascular and Endovascular Surgery, 2003
- Evaluation of 4-mm to 7-mm versus 6-mm prosthetic brachial-antecubital forearm loop access for hemodialysis: results of a randomized multicenter clinical trialJournal of Vascular Surgery, 2003
- Vascular access survival and incidence of revisions: A comparison of prosthetic grafts, simple autogenous fistulas, and venous transposition fistulas from the United States Renal Data System Dialysis Morbidity and Mortality StudyJournal of Vascular Surgery, 2001
- Permanent Vascular Access: A Nephrologist's ViewAmerican Journal of Kidney Diseases, 1993
- Long-term patency of polytetrafluoroethylene graft fistulasThe American Journal of Surgery, 1988