Hydrodynamic comparison of biological prostheses during progressive valve calcification in a simulated exercise situation. An in vitro study
Authors:
- Farhad BakhtiaryFBFarhad Bakhtiary
- Omer DzemaliOmer Dzemali
- Ulrich SteinseifferUSUlrich Steinseiffer
- Christof SchmitzCSChristof Schmitz
- Birgit GlasmacherBGBirgit Glasmacher
- Anton MoritzAMAnton Moritz
- Peter KleinePKPeter Kleine
- 1 November 2008
- journal article
- research article
- Published by Oxford University Press (OUP) in European Journal of Cardio-Thoracic Surgery
- Vol. 34 (5) , 960-963
Abstract
Despite continuous development of anticalcification treatment for biological valve prostheses, calcification remains one major cause of structural failure. The following study investigates hemodynamics and changes in opening and closing kinematics in progressively calcified porcine and pericardial valves in a simulated exercise situation. Five pericardial (Edwards Perimount Magna) and five porcine (Medtronic Mosaic Ultra) aortic valve bioprostheses (23 mm) were investigated in an artificial circulation system (150 beats/min, cardiac output 8l/min). Leaflet kinematics were visualized with a high-speed camera (3000 frames/s). Valves were exposed to a calcifying solution for 6 weeks. Repeated testing was performed every week. All prostheses underwent X-ray and photographic examination including measurement of calcium content for evaluation of progressive calcification. In the exercise situation pericardial valves demonstrated lower pressure gradients initially compared to the porcine valves (8.5+/-1.4 vs 11+/-1.6 mmHg), but significantly higher closing volume (5.3+/-1.2 ml vs 1.2+/-0.2 ml of stroke volume) leading to an equal total energy. Neither valve type demonstrated a significant increase in gradient or closing volume compared to the normal output situation. Opening and closing times were longer for pericardial valves after 6 weeks (opening time 42+/-10 ms vs 28+/-10 ms, closing time 84+/-12 vs 52+/-10 ms after 6 weeks). Pericardial valves calcified faster and more severely leading to an increase in gradients and closure volume. In the exercise situation pericardial valves demonstrated superior systolic function compared to porcine valves. Therefore pericardial valves have some advantage in active patients due to the lower gradients. Total energy loss remained constant during progressive calcification for both valves. Leaflet opening and closing is faster in porcine valves; clinical impact of these findings is not known. Diastolic performance is also important and should always be tested also in vivo.Keywords
This publication has 15 references indexed in Scilit:
- Opening and closing kinematics of fresh and calcified aortic valve prostheses: An in vitro studyThe Journal of Thoracic and Cardiovascular Surgery, 2007
- The Editor’s Roundtable: Cardiac Valve SurgeryThe American Journal of Cardiology, 2007
- Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten YearsThe Annals of Thoracic Surgery, 2007
- One year hemodynamic performance of the Perimount Magna pericardial xenograft and the Medtronic Mosaic bioprosthesis in the aortic position: a prospective randomized studyInteractive CardioVascular and Thoracic Surgery, 2007
- Impact of Patient-Prosthesis Mismatch and Aortic Valve Design on Coronary Flow Reserve After Aortic Valve ReplacementJACC, 2007
- Opening and Closure Characteristics of Different Types of Stented Biological ValvesThe Thoracic and Cardiovascular Surgeon, 2006
- Durability of pericardial versus porcine aortic valvesJACC, 2004
- Relation between size of prosthesis and valve gradient: comparison of two aortic bioprosthesisEuropean Journal of Cardio-Thoracic Surgery, 2004
- Mid-Term Hemodynamic and Clinical Results of the Stented Porcine Medtronic Mosaic Valve in Aortic PositionThe Thoracic and Cardiovascular Surgeon, 2004
- A four-dimensional study of the aortic root dynamicsEuropean Journal of Cardio-Thoracic Surgery, 2002