Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
Authors:
- Haya AlfarhanHAHaya Alfarhan
- Al- OtaibiAOAl- Otaibi
Open Access
- 1 July 2012
- journal article
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. 6, 1037-1043
Abstract
Purpose: To compare the precision of central corneal thickness (CCT) measurements taken with the handheld ultrasound pachymeter (USP), ultrasound biomicroscopy (UBM), and the Artemis-2 very high frequency ultrasound scanner (VHFUS) on normal subjects. Design: Prospective study. Methods: One eye from each of 61 normal subjects was randomly selected for this study. The measurements of the CCT were taken with the USP, VHFUS, and UBM. Results were compared statistically using repeated-measures analysis of variance (ANOVA), Pearson’s correlation coefficient, and limits of agreement. Results: The average CCT (± standard deviation) was 530.1 ± 30.5 µm, 554.9 ± 31.7 µm, and 559.5 ± 30.7 µm for UBM, VHFUS, and USP respectively. The intraobserver repeatability analyses of variance are not significant for USP, UBM, and VHFUS. P-values were 0.17, 0.19, and 0.37 respectively. Repeated-measures ANOVA showed a significant difference between the three different methods of measuring CCT (P = 0.0001). The ANOVA test revealed no statistically significant difference between USP and VHFUS (P > 0.05), yet statistical significant differences with UBM versus USP and UBM versus VHFUS (P < 0.001). There were high correlations between the three instruments (P < 0.0001). The mean differences (and upper/lower limits of agreement) for CCT measurements were 29.4 ± 14.3 (2.7/56), 4.6 ± 8.6 (-14.7/23.8), and -24.8 ± 13.1 (-50.4/0.8) for USP versus UBM, USP versus VHFUS, and UBM versus VHFUS, respectively. Conclusion: The UBM produces CCT measurements that vary significantly from those returned by the USP and the VHFUS, suggesting that the UBM may not be used interchangeably with either equipment for monitoring the CCT in the clinical setting.Keywords
This publication has 50 references indexed in Scilit:
- Intraoperator repeatability and interoperator reproducibility in the ultrasonic pachymetry measurements of central corneal thickness.2009
- Arc-scanning Very High-frequency Digital Ultrasound for 3D Pachymetric Mapping of the Corneal Epithelium and Stroma in Laser in situ KeratomileusisJournal of Refractive Surgery, 2000
- Corneal Thickness Measurements With the Topcon SP-2000P Specular Microscope and an Ultrasound PachymeterArchives of Ophthalmology (1950), 1999
- Ultrasound biomicroscopy. III. Accuracy and agreement of measurementsActa Ophthalmologica Scandinavica, 1999
- Evaluation of Central and Peripheral Corneal Thickness with Ultrasound Biomicroscopy in Normal and Keratoconic EyesCornea, 1997
- Epithelial and corneal thickness measurements by in vivo confocal microscopy through focusing (CMTF)Current Eye Research, 1997
- High-Frequency Ultrasound Measurement of the Thickness of the Corneal EpitheliumJournal of Refractive Surgery, 1993
- Reliability Coefficients of Three Corneal PachymetersAmerican Journal of Ophthalmology, 1992
- Evaluation and Comparison of Sources of Variability in the Measurement of Corneal Thickness with Ultrasonic and Optical PachymetersOphthalmic Surgery, Lasers and Imaging Retina, 1983
- Overnight swelling of the living human cornea.1980