Abstract
In analytical sciences, there is a general need for quality control to assess whether a product or a process meets defined requirements. Especially in proteomics, which implies analysis of ten thousands of analytes within a complex mixture, quality control to validate LC‐MS performance and method setup is inevitable to achieve day‐to‐day‐, inter‐system‐, as well as inter‐user reproducibility. Thus, results deriving from LC‐MS analyses can be benchmarked and the need for system maintenance can be revealed. In particular with the advent of label‐free quantification of peptides and proteins, which above all depends on highly stable and reproducible LC separations, HPLC performance has to be appropriately monitored throughout the entire analytical procedure to assure quality and validity of the obtained data. Oftentimes, proteolytic digests of standard proteins are used in this context; however, this approach implies some limitations, such as inadequate batch‐to‐batch reproducibility, limited (if any) dynamic range and compositional inflexibility. Here, we present an alternative strategy of nano‐LC‐MS/MS quality control based on a mixture of synthetic peptides covering the entire LC‐gradient as well as a dynamic range of more than two orders of magnitude. Thus, (i) reproducibility of LC separation, (ii) MS performance (including limit of detection, identification and quantification), as well as (iii) overall nano‐LC‐MS system performance and reproducibility can be routinely monitored even in highly complex samples.Keywords
Funding Information
- Ministerium für Innovation, Wissenschaft, Forschung und Technologie des Landes Nordrhein-Westfalen
- Bundesministerium für Forschung, Technologie und Raumfahrt
- MedSys project SARA (31P5800)
This publication has 21 references indexed in Scilit:
- Proteomics by Mass Spectrometry: Approaches, Advances, and ApplicationsAnnual Review of Biomedical Engineering, 2009
- Automated diagnosis of LC-MS/MS performanceBioinformatics, 2009
- Selected reaction monitoring for quantitative proteomics: a tutorialMolecular Systems Biology, 2008
- Phosphoproteome of Resting Human PlateletsJournal of Proteome Research, 2007
- Silver‐ and Coomassie‐staining protocols: Detection limits and compatibility with ESI MSElectrophoresis, 2007
- The absolute quantification strategy: a general procedure for the quantification of proteins and post-translational modificationsMethods, 2005
- Absolute quantification strategies in proteomics based on mass spectrometryExpert Review of Proteomics, 2004
- The proteome of Saccharomyces cerevisiae mitochondriaProceedings of the National Academy of Sciences of the United States of America, 2003
- Mass spectrometry-based proteomicsNature, 2003
- Anomalous cleavage of aspartyl-proline peptide bonds during amino acid sequence determinationsBiochemical and Biophysical Research Communications, 1970