Application of Chang's attenuation correction technique for single-photon emission computed tomography partial angle acquisition of Jaszczak phantom
Authors:
Open Access
- 1 January 2016
- journal article
- research article
- Published by Medknow in Journal of Medical Physics
- Vol. 41 (1) , 29-33
Abstract
The acquisition and processing of the Jaszczak phantom is a recommended test by the American College of Radiology for evaluation of gamma camera system performance. To produce the reconstructed phantom image for quality evaluation, attenuation correction is applied. The attenuation of counts originating from the center of the phantom is greater than that originating from the periphery of the phantom causing an artifactual appearance of inhomogeneity in the reconstructed image and complicating phantom evaluation. Chang's mathematical formulation is a common method of attenuation correction applied on most gamma cameras that do not require an external transmission source such as computed tomography, radionuclide sources installed within the gantry of the camera or a flood source. Tomographic acquisition can be obtained in two different acquisition modes for dual-detector gamma camera; one where the two detectors are at 180° configuration and acquire projection images for a full 360°, and the other where the two detectors are positioned at a 90° configuration and acquire projections for only 180°. Though Chang's attenuation correction method has been used for 360° angle acquisition, its applicability for 180° angle acquisition remains a question with one vendor's camera software producing artifacts in the images. This work investigates whether Chang's attenuation correction technique can be applied to both acquisition modes by the development of a Chang's formulation-based algorithm that is applicable to both modes. Assessment of attenuation correction performance by phantom uniformity analysis illustrates improved uniformity with the proposed algorithm (22.6%) compared to the camera software (57.6%).Keywords
This publication has 8 references indexed in Scilit:
- NIH Image to ImageJ: 25 years of image analysisNature Methods, 2012
- A comparison of 180 degrees and 360 degrees acquisition for attenuation-compensated thallium-201 SPECT images.1998
- Clinical evaluation of 360 degrees and 180 degrees data sampling techniques for transaxial SPECT thallium-201 myocardial perfusion imaging.1985
- Attenuation Correction for SPECT: An Evaluation of Hybrid ApproachesIEEE Transactions on Medical Imaging, 1984
- Tc‐99m attenuation coefficients in water‐filled phantoms determined with gamma camerasMedical Physics, 1984
- COMPARISON OF 180-DEGREES AND 360-DEGREES DATA-COLLECTION IN TL-201 IMAGING USING SINGLE-PHOTON EMISSION COMPUTERIZED-TOMOGRAPHY (SPECT) - CONCISE COMMUNICATION1982
- COMPARATIVE-STUDY OF THALLIUM EMISSION MYOCARDIAL TOMOGRAPHY WITH 180-DEGREES AND 360-DEGREES DATA-COLLECTION1982
- QUANTITATIVE MULTI-DETECTOR EMISSION COMPUTERIZED-TOMOGRAPHY USING ITERATIVE ATTENUATION COMPENSATION1982