Reply to comment on ‘origins of the changing detector response in small megavoltage photon radiation fields’
- 1 October 2018
- journal article
- editorial
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 63 (19) , 198002
Abstract
Andreo and Benmakhlouf (2017 Phys. Med. Biol . 62 1518–32) have disputed a finding of Scott et al (2012 Phys. Med. Biol . 57 4461–76) that the variation with field-size of the response of small ion chambers and solid-state dosimeters in small megavoltage photon radiation fields can largely be attributed to density. Further evidence for this finding was provided by Fenwick et al (2018 Phys. Med. Biol . 63 125003), but Andreo and Benmakhlouf (2018 Phys. Med. Biol . 63 125003) have now challenged the methodology used in that study. Specifically, Andreo and Benmakhlouf suggest that mass stopping-powers of fictitious materials used in Monte Carlo radiation transport calculations should be adjusted with material density according to the polarization effect, as if the materials were real and created by compressing other real materials. In this reply, we observe that fictitious materials are not real, and therefore their densities, mass stopping-powers and microscopic radiation interaction cross-sections can be freely and independently chosen to provide the clearest answers to the questions being studied. And we note that the key role played by density in small field detector response was further confirmed by our group back in 2013, using fictitious materials in which mass stopping-powers were adjusted with density, as preferred by Andreo and Benmakhlouf, as well as being held fixed, with very similar results being obtained in both circumstances (Underwood et al 2013a Med. Phys . 40 082102).Keywords
This publication has 16 references indexed in Scilit:
- Mass-density compensation can improve the performance of a range of different detectors under non-equilibrium conditionsPhysics in Medicine & Biology, 2013
- Detector density and small field dosimetry: Integral versus point dose measurement schemesMedical Physics, 2013
- Using cavity theory to describe the dependence on detector density of dosimeter response in non-equilibrium small fieldsPhysics in Medicine & Biology, 2013
- Characterizing the influence of detector density on dosimeter response in non-equilibrium small photon fieldsPhysics in Medicine & Biology, 2012
- Ionization chamber gradient effects in nonstandard beam configurationsMedical Physics, 2009
- Monte Carlo modeling of small photon fields: Quantifying the impact of focal spot size on source occlusion and output factors, and exploring miniphantom design for small‐field measurementsMedical Physics, 2009
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008
- Using a Monte Carlo model to predict dosimetric properties of small radiotherapy photon fieldsMedical Physics, 2008
- Density effect for the ionization loss of charged particles in various substancesAtomic Data and Nuclear Data Tables, 1984
- A General Theory of Cavity IonisationBritish Journal of Radiology, 1966