Purpose Ongoing discussions occur to translate the safety restrictions on MR scanners from specific absorption rate (SAR) to thermal dose. Therefore, this research focuses on the accuracy of thermal simulations in human subjects during an MR exam, which is fundamental information in that debate. Methods Radiofrequency (RF) heating experiments were performed on the calves of 13 healthy subjects using a dedicated transmit-receive coil while monitoring the temperature with proton resonance frequency shift (PRFS) thermometry. Subject-specific models and one generic model were used for electromagnetic and thermal simulations using Pennes' bioheat equation, with the blood equilibration constant equaling zero. The simulations were subsequently compared with the experimental results. Results The mean equaled 15 µT in the center slice of all volunteers, and 95% of the voxels had errors smaller than 2.8 µT between the simulation and measurement. The intersubject variation in RF power to achieve the required was 11%. The resulting intersubject variation in median temperature rise was 14%. Thermal simulations underestimated the median temperature increase on average, with 34% in subject-specific models and 28% in the generic model. Conclusions Although thermal measures are directly coupled to tissue damage and therefore suitable for RF safety assessment, insecurities in the applied thermal modeling limit their estimation accuracy. Magn Reson Med, 2016.Keywords
This publication has 30 references indexed in Scilit:
- Rapid method for thermal dose‐based safety supervision during MR scansBioelectromagnetics, 2015
- In vivo radiofrequency heating in swine in a 3T (123.2-MHz) birdcage whole body coilMagnetic Resonance in Medicine, 2013
- CEM43°C thermal dose thresholds: a potential guide for magnetic resonance radiofrequency exposure levels?European Radiology, 2013
- Thermal Tissue Damage Model Analyzed for Different Whole‐Body SAR and Scan Durations for Standard MR Body CoilsMagnetic Resonance in Medicine, 2013
- MEDICAL MAGNETIC RESONANCE (MR) PROCEDURES: PROTECTION OF PATIENTSHealth Physics, 2004
- A description of discrete vessel segments in thermal modelling of tissuesPhysics in Medicine & Biology, 1996
- Recent developments in modeling heat transfer in blood perfused tissuesIEEE Transactions on Biomedical Engineering, 1994
- Thermal effects of MR imaging: worst-case studies on sheep.American Journal of Roentgenology, 1990
- Superficial- and deep-tissue temperature increases in anesthetized dogs during exposure to high specific absorption rates in a 1.5-T MR imager.Radiology, 1988
- Thermal dose determination in cancer therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1984