A low complexity, low power, programmable QRS detector based on wavelet transform for Implantable Pacemaker IC
- 1 September 2006
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21641676,p. 160-163
Abstract
This paper presents a QRS detector operating in multi-mode with low complexity, power saving consumption for implantable pacemaker IC. The complexity of detector is optimized after testing dynamic range of internal signals. Correspondingly, static power consumption is reduced significantly. The dynamic power consumption is saved by using down-sampling rate structure and turning off de- active filter banks bases on noise estimation. This QRS detector is evaluated with MIT-BIH Arrhythmia database archives >99% detection ratio. A soft IP core is prototyped on FPGA and synthesized by Samsung 0.18um standard library.Keywords
This publication has 7 references indexed in Scilit:
- Wavelet transforms and the ECG: a reviewPhysiological Measurement, 2005
- A very low-power CMOS mixed-signal IC for implantable pacemaker applicationsIEEE Journal of Solid-State Circuits, 2004
- An Invariance property of the generalized likelihood ratio testIEEE Signal Processing Letters, 2003
- The principles of software QRS detectionIEEE Engineering in Medicine and Biology Magazine, 2002
- Optimal and wavelet-based shock wave detection and estimationThe Journal of the Acoustical Society of America, 1998
- A comparison of the noise sensitivity of nine QRS detection algorithmsIEEE Transactions on Biomedical Engineering, 1990
- A theory for multiresolution signal decomposition: the wavelet representationIEEE Transactions on Pattern Analysis and Machine Intelligence, 1989