A selective ventilation distribution circuit (S.V.D.C.)
Authors:
- 1 June 1979
- journal article
- case report
- Published by Springer Nature in Intensive Care Medicine
- Vol. 5 (2) , 95-98
Abstract
A circuit for selective ventilation distribution (S. V.D.C.) is described. The use of the circuit is illustrated with five cases in which it was used S.V.D.C. allowed: 1) Measurement of individual lung tidal volumes and pressures during the respiratory cycle. 2) Adjustment of the tidal volumes of both lungs in the most appropriate way for the patient. 3) Application of different degrees of PEEP to each lung. The circuit is easily built and managed, the patient is connected to it by a double lumen endotracheal tube (Carlens tube); the circuit requires only one respirator.Keywords
This publication has 11 references indexed in Scilit:
- Differential lung ventilation with PEEP in the treatment of unilateral pneumoniaCritical Care Medicine, 1977
- Therapy of unilateral pulmonary insufficiency with a double lumen endotracheal tubeCritical Care Medicine, 1976
- Differential ventilation of the lungs in manCritical Care Medicine, 1976
- Correlation of positive end-expiratory pressure with cardiovascular performanceCritical Care Medicine, 1975
- Physiologic Consequences of Positive End-Expiratory Pressure (PEEP) VentilationAnnals of Surgery, 1973
- Physiologic shunting in the lung in critically ill or injured patientsJournal of Surgical Research, 1970
- Continuous Positive-Pressure Ventilation in Acute Respiratory FailureNew England Journal of Medicine, 1970
- The Mechanism of Arterial Hypoxemia in Acute Myocardial InfarctionCirculation, 1970
- Continuous positive-pressure breathing (CPPB) in adult respiratory distress syndromeThe Journal of Thoracic and Cardiovascular Surgery, 1969
- PHYSIOLOGICAL STUDIES OF THE EFFECTS OF INTERMITTENT POSITIVE PRESSURE BREATHING ON CARDIAC OUTPUT IN MANAmerican Journal of Physiology-Legacy Content, 1947