Characteristics of the electron beam outside the applicator in an Elekta Versa HD Linac
Open Access
- 18 March 2025
- journal article
- research article
- Published by Springer Nature in Physical and Engineering Sciences in Medicine
- Vol. 48 (2) , 635-647
Abstract
Radiotherapy is an essential component of cancer treatment, but healthy tissues can be exposed to out-of-field doses, potentially causing adverse effects and secondary cancers. This study investigates peripheral doses outside the electron beam applicator in an Elekta Versa HD linear accelerator. Peripheral doses outside an electron applicator were measured using 6, 9, and 12 MeV beams at their respective maximum dose depths while maintaining a 100 cm source-to-surface distance. Measurements employed EBT3 films within Plastic Water DT phantoms. The influence of field size on penumbra width and peripheral doses were examined using various cutouts (6 × 6 cm², 10 × 10 cm², and a 5 cm diameter circle) within a 10 × 10 cm² applicator, with gantry and collimator angles set to 0 degrees. Additionally, the impact of collimator angles on penumbra width and peripheral doses was explored, enhancing the understanding of dose distribution. Measured profiles were also compared with those calculated using Monaco treatment planning system. Findings showed that both penumbra width and peripheral dose values increased with energy across different field sizes and collimator angles. Root Mean Square Deviation (RMSD) analysis indicated deviations of 1.8 mm for penumbra and 1.1% for peripheral doses between measured profiles and Treatment Planning System (TPS) predictions for all field sizes. Peripheral doses remained below 5% of the maximum dose at distances ranging from 10 to 15 mm away from the field edges, indicating acceptable tolerance levels (ICRU report 24). However, further dose reduction may be possible with additional shielding to keep doses as low as reasonably achievable. This study highlights the critical importance of considering peripheral doses in radiotherapy, emphasizing the need to evaluate the impact on healthy tissues outside the primary treatment area to ensure patient safety and mitigate long-term treatment-related side effects. The findings underscore the necessity of implementing appropriate measures to minimize peripheral doses.Keywords
All Related Versions
Funding Information
- University of Western Australia
This publication has 31 references indexed in Scilit:
- Out of field scatter from electron applicator in modern linear acceleratorsJournal of Applied Clinical Medical Physics, 2024
- A review on radiochromic film dosimetry for dose verification in high energy photon beamsRadiation Physics and Chemistry, 2021
- Shielding effect of a lead apron on the peripheral radiation dose outside the applicator of electron beams from an Elekta linear acceleratorJournal of Applied Clinical Medical Physics, 2020
- Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG‐55Medical Physics, 2020
- The peripheral dose outside the applicator in electron beams of an Elekta linear acceleratorAustralasian Physics & Engineering Sciences in Medicine, 2018
- Neutron production during the interaction of monoenergetic electrons with a Tungsten foil in the radiotherapeutic energy rangeNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2017
- Investigating in-field and out-of-field neutron contamination in high-energy medical linear accelerators based on the treatment factors of field size, depth, beam modifiers, and beam typePhysica Medica, 2015
- Monte Carlo simulation and measurement of radiation leakage from applicators used in external electron radiotherapyPhysica Medica, 2013
- ICRP PUBLICATION 118: ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs — Threshold Doses for Tissue Reactions in a Radiation Protection ContextAnnals of the ICRP, 2012
- Radiation leakage from linac electron applicator assemblyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1984