Determination of Dose Rates from Homogeneously Distributed Beta,Gamma- and EC,Gamma-emitters in Phantom Organs Using a Liquid Scintillator

Authors:
  • KTK. Tagder
    K. Tagder
  • ISI. Schmitz-Feuerhake
    I. Schmitz-Feuerhake

Abstract
Dose rates from radioactivity distributed homogeneously in simulated organs with volumes up to 310 ml have been investigated using a liquid scintillator. From the pulse-height distributions, the mean absorbed energy per decay has been estimated and compared to theoretical dose-rate values. Several nuclides of interest in nuclear medicine - 51Cr, 75Se, 82Br, 85Sr, 131I, 197Hg,198Au and 203Hg - were measured. The maximum error in the experimental determination is estimated to be within 5% in most cases. Deviations up to 35% were found in comparisons with the theoretical values.