Correcting for motion artifact in handheld laser speckle images
- 15 March 2018
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Biomedical Optics
- Vol. 23 (03) , 036006-7
Abstract
Laser speckle imaging (LSI) is a wide-field optical technique that enables superficial blood flow quantification. LSI is normally performed in a mounted configuration to decrease the likelihood of motion artifact. However, mounted LSI systems are cumbersome and difficult to transport quickly in a clinical setting for which portability is essential in providing bedside patient care. To address this issue, we created a handheld LSI device using scientific grade components. To account for motion artifact of the LSI device used in a handheld setup, we incorporated a fiducial marker (FM) into our imaging protocol and determined the difference between highest and lowest speckle contrast values for the FM within each data set (Kbest and Kworst). The difference between Kbest and Kworst in mounted and handheld setups was 8% and 52%, respectively, thereby reinforcing the need for motion artifact quantification. When using a threshold FM speckle contrast value (KFM) to identify a subset of images with an acceptable level of motion artifact, mounted and handheld LSI measurements of speckle contrast of a flow region (KFLOW) in in vitro flow phantom experiments differed by 8%. Without the use of the FM, mounted and handheld KFLOW values differed by 20%. To further validate our handheld LSI device, we compared mounted and handheld data from an in vivo porcine burn model of superficial and full thickness burns. The speckle contrast within the burn region (KBURN) of the mounted and handheld LSI data differed by <4 % when accounting for motion artifact using the FM, which is less than the speckle contrast difference between superficial and full thickness burns. Collectively, our results suggest the potential of handheld LSI with an FM as a suitable alternative to mounted LSI, especially in challenging clinical settings with space limitations such as the intensive care unit.Keywords
Funding Information
- Arnold and Mabel Beckman Foundation
- Air Force Office of Scientific Research (FA9550-17-1-0193)
- US Air Force Office of Scientific Research (FA9550-17-1-0193)
- National Institutes of Health (P41 EB015890)
- National Institutes of Health (P41 EB015890)
- National Institute of Health Laser Microbeam Medical Program (P41 EB015890)
- National Institutes of Health Laser Microbeam Medical Program (P41 EB015890)
- National Science Foundation (DGE-1321846)
- NSF GRFP (DGE-1321846)
- National Science Foundation (DGE-1321846)
- Air Force Office of Scientific Research (FA9550-14-00340)
- Air Force Office of Scientific Research (FA9550-16-1-0342)
- National Institute of Biomedical Imaging and Bioengineering (R01GM108634)
- National Institutes of Health (R01GM108634)
- NIH/NIGMS (R01GM108634)
- National Institutes of Health (TL1 TR001415)
- National Institutes of Health (5T32HL116270)
This publication has 20 references indexed in Scilit:
- Intraoperative, real‐time monitoring of blood flow dynamics associated with laser surgery of port wine stain birthmarksLasers in Surgery and Medicine, 2015
- Optimisation of movement detection and artifact removal during laser speckle contrast imagingMicrovascular Research, 2015
- Talaporfin Sodium–Mediated Photodynamic Therapy Alone and in Combination with Pulsed Dye Laser on Cutaneous VasculatureJournal of Investigative Dermatology, 2015
- Quantitative assessment of graded burn wounds in a porcine model using spatial frequency domain imaging (SFDI) and laser speckle imaging (LSI)Biomedical Optics Express, 2014
- Longitudinal In Vivo Imaging to Assess Blood Flow and Oxygenation in Implantable Engineered TissuesTissue Engineering Part C: Methods, 2012
- Preclinical in vivo evaluation of Npe6‐mediated photodynamic therapy on normal vasculatureLasers in Surgery and Medicine, 2012
- Wide-field functional imaging of blood flow and hemoglobin oxygen saturation in the rodent dorsal window chamberMicrovascular Research, 2011
- Laser speckle contrast imaging accurately measures blood flow over moving skin surfacesMicrovascular Research, 2011
- Microvascular flow, clinical illness severity and cardiovascular function in the preterm infantArchives of Disease in Childhood - Fetal and Neonatal Edition, 2008
- Assessment of burn depth: A prospective, blinded comparison of laser Doppler imaging and videomicroscopyBurns, 2007