Computed Tomography Thermography for Ablation Zone Prediction in Microwave Ablation and Cryoablation: Advantages and Challenges in an Ex Vivo Porcine Liver Model
- 1 September 2020
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 44 (5) , 744-749
Abstract
Purpose The aim of this study was to investigate the diagnostic accuracy of computed tomography (CT) for the prediction of ablation zones from microwave ablation (MWA) and cryoablation (CA) in an ex vivo porcine liver model. Methods Sequential (30 seconds) CT scans were acquired during and after MWA and CA in an ex vivo porcine liver model. We generated 120-kVp equivalent reconstructions of generic dual-energy CT data sets, and comprehensive region-of-interest measurements were statistically correlated with invasive temperature monitoring using Pearson correlation coefficient. Binary logistic regression was performed for prediction of successful ablation. Results With the use of pooled data from 6 lesions in 2 separate experiments, correlation analysis of attenuation in Hounsfield units (HU) and temperature yieldedr= -0.79 [confidence interval (CI), -0.85 to -0.71] for MWA andr= 0.62 (CI, 0.55 to 0.67) for CA. For MWA, there was a linear association between attenuation and temperature up to 75 degrees C; thus, linear regression yielded a slope of -2.00 HU/degrees C (95% CI, -1.58 to -2.41). For CA, a linear association between attenuation and temperature was observed in the cooling phase with a slope of 2.11 HU/degrees C (95% CI, 1.79 to 2.58). In MWA treatment, binary logistic regression separated less than 70 degrees C and greater than 70 degrees C with 89.2% accuracy. Within the ice ball, temperatures above and below -20 degrees C were distinguished with 65.3% accuracy. Conclusions Our experiments reveal several difficulties in predicting ablation zone temperature from CT attenuation. Microwave ablation leads to gas production in the tissue, which degrades the accuracy of noninvasive temperature measurement, especially at higher temperatures. In CA, CT thermometry is limited by ice ball formation, which leads to homogeneous attenuation, nearly independent of temperature. Further research is needed to define the role of CT thermography in ablation zone monitoring in liver malignancies.Keywords
This publication has 27 references indexed in Scilit:
- Instant Outcome Evaluation of Microwave Ablation With Subtraction CT in an In Vivo Porcine ModelInvestigative Radiology, 2019
- Ablation treatment of primary and secondary liver tumors under contrast-enhanced ultrasound guidance in field practice of interventional ultrasound centers. A multicenter studyEuropean Journal of Radiology, 2018
- Computed Tomography–Guided Tumor AblationJournal of Computer Assisted Tomography, 2017
- Real-Time Magnetic Resonance Imaging–Guided Cryoablation of Small Renal Tumors at 1.5 TInvestigative Radiology, 2013
- Assessment of liver tumor response by high-field (3T) MRI after radiofrequency ablation: Short- and mid-term evolution of diffusion parameters within the ablation zoneEuropean Journal of Radiology, 2012
- Pre-clinical testing of a phased array ultrasound system for MRI-guided noninvasive surgery of the brain—A primate studyEuropean Journal of Radiology, 2006
- Non‐invasive thermometry for monitoring hepatic radiofrequency ablationMinimally Invasive Therapy & Allied Technologies, 2006
- Monitoring Tissue Coagulation During Thermoablative Treatment by Using a Novel Magnetic Resonance Imaging Contrast AgentInvestigative Radiology, 2004
- Noninvasive thermometry with a clinical x‐ray CT scannerMedical Physics, 1982
- The Temperature Dependence of Computed Tomography Attenuation ValuesJournal of Computer Assisted Tomography, 1979