While the titrimetric assay is one of the most precise analytical techniques available, only a limited list of complexometric chelators is available, as many otherwise promising reagents are not water-soluble. Recent work demonstrated successful titrimetry with ion-exchanging polymeric nanospheres containing hydrophobic complexing agents, so-called ionophores, opening an exciting avenue in this field. However, this method was limited to ionophores of very high affinity to the analyte and exhibited a relatively limited titration capacity. To overcome these two limitations, we report here on solvent based titration reagents. This heterogeneous titration principle is based on the dissolution of all hydrophobic recognition components in a solvent such as dichloromethane (CH2Cl2) where the ionophores are shown to maintain a high affinity to the target ions. HSV (hue, saturation, value) analysis of the images captured with a digital camera provides a convenient and inexpensive way to determine the end point. This approach is combined with an automated titration setup. The titrations of the alkali metals K+, Na+, and Li+ in aqueous solution are successfully demonstrated. The potassium concentration in human serum without pretreatment was precisely and accurately determined as 4.38 mM ± 0.10 mM (automated titration), which compares favorably with atomic emission spectroscopy (4.47 mM ± 0.20 mM).Keywords
Funding Information
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
- Université de Genève
- China Scholarship Council
This publication has 36 references indexed in Scilit:
- Direct Fluorescent Detection of Blood Potassium by Ion-Selective Formation of Intermolecular G-Quadruplex and Ligand BindingAnalytical Chemistry, 2016
- Specific detection of potassium ion in serum by a modified G-quadruplex methodRSC Advances, 2016
- Understanding Complexometric Titrations of Metal Cations with Aminopolycarboxylic Acids (EDTA and Analogs) within the frame of the Notion of Reactions between Groups of Chemical SpeciesWorld Journal of Chemical Education, 2015
- Analysis of Human Serum and Whole Blood for Mineral Content by ICP-MS and ICP-OES: Development of a Mineralomics MethodBiological Trace Element Research, 2014
- A Defect in the Kv Channel-Interacting Protein 2 (KChIP2) Gene Leads to a Complete Loss of Ito and Confers Susceptibility to Ventricular TachycardiaCell, 2001
- Ion homeostasis and apoptosisCurrent Opinion in Cell Biology, 2001
- Detection techniques in ion chromatography of inorganic ionsTrAC Trends in Analytical Chemistry, 2001
- Increased Blood Pressure during Potassium Depletion in Normotensive MenNew England Journal of Medicine, 1989
- Studies on bis(crown ether)-based ion-selective electrodes for the potentiometric determination of sodium and potassium in serumThe Analyst, 1989
- Flame spectrometric determination of sodium, potassium and calcium in blood serum by measurement of microsamplesMicrochimica Acta, 1973