Falsely Increased Blood Tacrolimus Concentrations Using the ACMIA Assay Due to Circulating Endogenous Antibodies in a Liver Transplant Recipient: A Tentative Approach to Obtaining Reliable Results
Authors:
- J Carlos TutorJTJ Carlos Tutor
- 1 April 2009
- journal article
- case report
- Published by Wolters Kluwer Health in Therapeutic Drug Monitoring
- Vol. 31 (2) , 269-272
Abstract
The presence of endogenous antibodies in the serum of some patients has long been known to be a potential source of interference in immunoassays. We report falsely increased whole blood tacrolimus concentrations using the antibody conjugated magnetic immunoassay from Siemens HealthCare Diagnostics in a liver transplant recipient due to the presence of circulating endogenous antibodies (possibly heterophilic antibodies). Estimation of whole blood tacrolimus concentrations from the washed erythrocytes concentrations is proposed as a tentative approach for obtaining reliable results using the antibody conjugated magnetic immunoassay assay in these cases, leading to analogous blood tacrolimus concentrations to those produced by the microparticle enzyme immunoassay from Abbott Laboratories.Keywords
This publication has 18 references indexed in Scilit:
- Comparison of Blood Tacrolimus Concentrations in Liver and Kidney Transplant Recipients using ACMIA and MEIA ImmunoassaysUpsala Journal of Medical Sciences, 2008
- Erroneous laboratory results: what clinicians need to knowClinical Medicine, 2007
- On detecting interference from endogenous antibodies in immunoassays by doubling dilutions testClinical Chemistry and Laboratory Medicine (CCLM), 2007
- A Radical Approach Is Needed to Eliminate Interference from Endogenous Antibodies in ImmunoassaysClinical Chemistry, 2005
- Towards a better understanding of heterophile (and the like) antibody interference with modern immunoassaysClinica Chimica Acta, 2002
- Interference in immunoassay is an underestimated problemAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2002
- Bioanalytical Method Validation—A Revisit with a Decade of ProgressPharmaceutical Research, 2000
- Therapeutic drug monitoring of cyclosporine and tacrolimusClinical Biochemistry, 1998
- Analytical methods validation: Bioavailability, bioequivalence and pharmacokinetic studiesEuropean Journal of Drug Metabolism and Pharmacokinetics, 1991
- The Determination of Calcium, Glucose, Urea and Uric Acid Using the Kodak EKTACHEM Multilayer Film Technology: An EvaluationClinical Chemistry and Laboratory Medicine (CCLM), 1982