Improved large tip angle parallel transmission pulse design through a perturbation analysis of the bloch equation
Open Access
- 21 April 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 66 (3) , 687-696
Abstract
Parallel transmission has emerged as an efficient means for implementing multidimensional spatially selective radiofrequency excitation pulses. To date, most theoretical and experimental work on parallel transmission radiofrequency (RF) pulse design is based on the small‐tip‐angle approximation to the Bloch equation. The small‐tip‐angle, while mathematically compact, is not an exact solution and leads to significant errors when large‐tip‐angle pulses are designed. Methods have been proposed to overcome the limitations of the small‐tip‐angle using regularized least‐square optimization or optimal control algorithms. These methods, however, are based on further approximations to the Bloch equation or require the use of general purpose algorithms that do not capitalize fully on the dynamics of the physical model at hand. In this article, a novel algorithm for large‐tip‐angle parallel transmission pulse design is proposed. The algorithm relies on a perturbation analysis of the Bloch equation and it depicts the relationship between the excited magnetization, its deviation from the target pattern and the desired pulses. Simulations and experiments are used to validate the proposed method on a 7T 8‐channel transmit array. The results demonstrate that the perturbation analysis algorithm provides a fast and accurate approach for multidimensional large‐tip‐angle pulse design, especially when large acceleration factors and/or echo‐planar trajectories are used. Magn Reson Med, 2011.Keywords
Funding Information
- NIH (CA106840)
- Siemens Medical Systems
This publication has 22 references indexed in Scilit:
- Time‐optimal design for multidimensional and parallel transmit variable‐rate selective excitationMagnetic Resonance in Medicine, 2009
- Slice‐selective RF pulses for in vivo B inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16‐element coilMagnetic Resonance in Medicine, 2008
- Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channelsMagnetic Resonance in Medicine, 2008
- Designing multichannel, multidimensional, arbitrary flip angle RF pulses using an optimal control approachMagnetic Resonance in Medicine, 2008
- Saturated double‐angle method for rapid B1+ mappingMagnetic Resonance in Medicine, 2006
- Parallel imaging reconstruction using automatic regularizationMagnetic Resonance in Medicine, 2004
- Spatial localization in two dimensions using NMR designer pulsesJournal of Magnetic Resonance (1969), 1989
- A linear class of large-tip-angle selective excitation pulsesJournal of Magnetic Resonance (1969), 1989
- A k-space analysis of small-tip-angle excitationJournal of Magnetic Resonance (1969), 1989
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988