Non-intrusive PODI-ROM for patient-specific aortic blood flow in presence of a LVAD device
Authors:
- 11 July 2022
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 107 (1) , 103849
Abstract
Left ventricular assist devices (LVADs) are used to provide haemodynamic support to patients with critical cardiac failure. Severe complications can occur because of the modifications of the blood flow in the aortic region. In this work, the effect of a continuous flow LVAD device on the aortic flow is investigated by means of a non-intrusive reduced order model (ROM) built using the proper orthogonal decomposition with interpolation (PODI) method based on radial basis functions (RBF). The full order model (FOM) is represented by the incompressible Navier-Stokes equations discretized by using a Finite Volume (FV) technique, coupled with three-element Windkessel models to enforce outlet boundary conditions in a multi-scale approach. A patient-specific framework is proposed: a personalized geometry reconstructed from Computed Tomography (CT) images is used and the individualization of the coefficients of the three-element Windkessel models is based on experimental data provided by the Right Heart Catheterization (RHC) and Echocardiography (ECHO) tests. At FOM level, we also consider the pre-surgery configuration in order to further validate the predictive capabilities of the model in several contexts. The ROM has been tested by considering a parametric setting with respect to the LVAD flow, which is a crucial parameter of the problem. We consider a parameter range that covers typical clinical values. The accuracy of the ROM is assessed against results obtained with the FOM both for primal, velocity and pressure, and derived quantities, wall shear stress (WSS). Finally, we briefly discuss the efficiency of our ROM approach.Keywords
All Related Versions
This publication has 55 references indexed in Scilit:
- 4D flow magnetic resonance imaging in bicuspid aortic valve disease demonstrates altered distribution of aortic blood flow helicityMagnetic Resonance in Medicine, 2013
- Numerical simulation of left ventricular assist device implantations: Comparing the ascending and the descending aorta cannulationsMedical Engineering & Physics, 2013
- Importance of realistic LVAD profiles for assisted aortic simulations: evaluation of optimal outflow anastomosis locationsComputer Methods in Biomechanics and Biomedical Engineering, 2012
- High Fluid Shear Stress and Spatial Shear Stress Gradients Affect Endothelial Proliferation, Survival, and AlignmentAnnals of Biomedical Engineering, 2011
- Numerical Modeling of Hemodynamics with Pulsatile Impeller Pump SupportAnnals of Biomedical Engineering, 2010
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceThe New England journal of medicine, 2009
- Aortic valve thrombosis after implantation of temporary left ventricular assist deviceInteractive CardioVascular and Thoracic Surgery, 2009
- Effect of Left Ventricular Assist Device Outflow Conduit Anastomosis Location on Flow Patterns in the Native AortaASAIO Journal, 2006
- The History of Poiseuille's LawAnnual Review of Fluid Mechanics, 1993
- A calculation procedure for heat, mass and momentum transfer in three-dimensional parabolic flowsInternational Journal of Heat and Mass Transfer, 1972