Impact of modulating phase duration on electrically evoked auditory brainstem responses obtained during cochlear implantation
Authors:
- N-X BonneNBN-X Bonne
- D DouchementDDD Douchement
- J DesruellesJDJ Desruelles
- P FayouxPFP Fayoux
- I RuzzaIRI Ruzza
- C VincentCVC Vincent
- 28 August 2014
- journal article
- Published by Taylor & Francis in Cochlear Implants International
- Vol. 16 (3) , 168-174
Abstract
Objective To investigate the effect of increasing phase duration (pulse width, T-pulse) using a biphasic pulse composed of an initial anodic active phase followed by a balancing cathodic phase on the electrically evoked auditory brainstem responses (eABRs) recorded at the time of cochlear implantation. Design eABRs recorded during 188 surgeries for cochlear implantation from 1999 to 2006 in a single center were retrospectively reviewed by two independent observers. All patients were fitted with a NEURELEC cochlear implant (CI) device, initially DIGISONIC® then DIGISONIC SP® (2004–2006). Result Immediately following cochlear implantation, stimulation by the CI resulted in reliable wave III and V eABR waveforms (mean wave III latency 2.23 ± 0.38 ms SD and wave V latency 4.28 ± 0.42 ms SD). Latencies followed an apical to basal gradient (0.32 ms increase in mean eV latency and 0.12 ms for eIII latency). With increasing phase duration, wave III and wave V latencies significantly decreased in association with a shortening of the eIII–eV interwave gap, while amplitudes of both waves increased. Conclusion The impact of increasing phase duration on latency and amplitude of brainstem responses in a large set of patients implanted with NEURELEC CIs was reported.Keywords
This publication has 23 references indexed in Scilit:
- Effects of stimulus manipulation on electrophysiological responses in pediatric cochlear implant users. Part I: Duration effectsHearing Research, 2008
- Effects of stimulus manipulation on electrophysiological responses of pediatric cochlear implant users. Part II: Rate effectsHearing Research, 2008
- Higher Sensitivity of Human Auditory Nerve Fibers to Positive Electrical CurrentsJournal of the Association for Research in Otolaryngology, 2008
- Neurophysiology of Cochlear Implant Users II: Comparison Among Speech Perception, Dynamic Range, and Physiological MeasuresEar & Hearing, 2002
- Loudness Growth Functions and EABR Characteristics in Digisonic Cochlear ImplanteesActa Oto-Laryngologica, 1999
- Clinical results for the digisonic multichannel cochlear implantThe Laryngoscope, 1995
- Speech Coding Strategies of the Digisonic Fully Digitized Cochlear ImplantActa Oto-Laryngologica, 1995
- Effects of phase duration on detection of electrical stimulation of the human cochleaHearing Research, 1993
- Electrically evoked auditory brainstem response to stimulation of different sites in the cochleaHearing Research, 1993
- Characteristics of Electrically Evoked ‘Auditory’ Brainstem Responses Elicited with the Nucleus 22-Electrode Intracochlear ImplantScandinavian Audiology, 1990