Intracavitary dosimetry of a high-activity remote loading device with oscillating source

Authors:

Abstract
Dosimetric experiments were carried out in order to obtain the dose distribution in water around a Flectcher applicator loaded by a Buchler system containing two 137Cs 148 GBq [gigabecquerel] (4 Ci) sources and one 192Ir 740 GBq (20 Ci) source. The mechanical system which controls the movement of the 192Ir source and the resulting motion of the source are described. The dose distribution around the sources was measured photographically and by a PWT Normal 0.22 cm3 ionization chamber. The absolute dose rate was measured along the lateral axes of the sources. The measurements of exposure in water near the sources were corrected for the effect due to the finite volume of the chamber. The quantization method described by Cassell (1983) was utilized to calculate the variation of the dose rate along the lateral axes of the sources. The dose distribution around both 192Ir and 137Cs sources was spherical for angles greater than 40.degree. from the longitudinal axes of the sources. A simple algorithm fitting the data for the moving 192Ir source is proposed. A program written in FORTRAN IV and run on a Univac 1100/80 computer was used to plot dose distributions on anatomical data obtained from CT [computed tomography] images. The uncertainties of the measurements and calculations have been examined and the greatest error has been 5.5%. The clinical significance of the treatment method is discussed.

This publication has 14 references indexed in Scilit: