The Determination of Hemoglobin in Blood Banks
Authors:
- 2 January 1964
- journal article
- research article
- Published by Wiley in Transfusion
- Vol. 4 (1) , 45-49
Abstract
The accuracy of the copper sulfate method of determining hemoglobin was investigated under usual blood bank conditions. In our hands this method was found to be grossly inaccurate. Using the cyanmethemoglobin technic, it was demonstrated that 83.9 per cent of females and 80.4 per cent of males rejected as having a low hemoglobin by the copper sulfate method, actually had hemoglobin values above minimum requirements for blood donors. Detailed instruction as to technics and pitfalls of the copper sulfate method was ineffectual in correcting errors as encountered by a large nursing staff carrying out the procedure. Donors found to have acceptable hemoglobin values by the copper sulfate method appeared to be safely accepted as donors; cyanmethemoglobin duplicate checks did not reveal any discrepancies in this group.Keywords
This publication has 3 references indexed in Scilit:
- Standards for Rejection of Blood Donors: A Comparison of CuSO4 Specific Gravity, Microhematocrit and Electronic Hematocrit Values with Hemoglobin Values by the Cyanmethemoglobin TechnicTransfusion, 1962
- A Study of Donor Rejection Rates Due to Low Hemoglobin Levels1958
- MEASUREMENT OF SPECIFIC GRAVITIES OF WHOLE BLOOD AND PLASMA BY STANDARD COPPER SULFATE SOLUTIONSJournal of Biological Chemistry, 1950