Characterization of the radiation quality of60Co therapy units by the fraction of air kerma attributable to scattered photons
Authors:
- H K LooeHLH K Looe
- R P KapschRKR P Kapsch
- D HarderDHD Harder
- K C WillbornKWK C Willborn
- A RühmannARA Rühmann
- B PoppeBPB Poppe
- 16 March 2007
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (7) , N137-N147
Abstract
In this work we present a new parameter for characterizing the emitted photon spectra of (60)Co radiotherapy units. It is intended to propose this parameter for the revised DIN standard 6809-1. In the previous DIN regulation, it had been sufficient to state the nature of the radioactive material within the source. However, scatter processes within the radioactive material as well as the source housing and the collimator system influence the shape of the photon spectrum, with a noticeable contribution in the low-energy portion. The fraction of the air kerma for a given distance from the source, position and beam size in air comprising all contributions by scattered photons up to an upper energy limit for the emitted spectrum from (60)Co decay, will be proposed as a typical parameter. The new quantity, which is termed the 'fraction of air kerma attributable to scattered photons', P(E)(Scatter), has been calculated for E = 1.17 MeV and compared for four different Monte Carlo-simulated spectra of used (60)Co devices. Not included in this new formalism is the air kerma contribution by scattered photons in between the two lines of the (60)Co spectrum. A simple measurement procedure based on the signal ratio of two Farmer chamber detectors with different wall materials is discussed and its feasibility shown.Keywords
This publication has 14 references indexed in Scilit:
- Study of scattered photons from the collimator system of Leksell Gamma Knife using the EGS4 Monte Carlo CodeMedical Physics, 2005
- Compton‐scattering measurement of diagnostic x‐ray spectrum using high‐resolution Schottky CdTe detectorMedical Physics, 2005
- Spectra of scattered photons in large absorbers and their importance for the values of radiation weighting factor wRRadiation Protection Dosimetry, 2004
- Angular measurement of the Cobalt-60 emitted radiation spectrum from a radiosurgery irradiatorMedical Physics, 2001
- A Monte Carlo study of the quality dependence factors of common TLD materials in photon and electron beamsPhysics in Medicine & Biology, 1998
- Application of cavity theory to the response of various TLDs to gammas degraded in waterPhysics in Medicine & Biology, 1997
- Comparison of x‐ray cross sections for diagnostic and therapeutic medical physicsMedical Physics, 1996
- The quality dependence of LiF TLD in megavoltage photon beams: Monte Carlo simulation and experimentsPhysics in Medicine & Biology, 1996
- Measurement of accelerator bremsstrahlung spectra with a high‐efficiency Ge detectorMedical Physics, 1991
- Spectral Distribution of Scattered Radiation from a Kilocurie Cobalt 60 UnitBritish Journal of Radiology, 1958