Design, simulation and evaluation of kinematic alternatives for Insertable Robotic Effectors Platforms in Single Port Access Surgery
Authors:
- 1 May 2010
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10504729,p. 1053-1058
Abstract
This paper presents the task specifications for designing a novel Insertable Robotic Effectors Platform (IREP) with integrated stereo vision and surgical intervention tools for Single Port Access Surgery (SPAS). This design provides a compact deployable mechanical architecture that may be inserted through a single Ø15 mm access port. Dexterous surgical intervention and stereo vision are achieved via the use of two snake-like continuum robots and two controllable CCD cameras. Simulations and dexterity evaluation of our proposed design are compared to several design alternatives with different kinematic arrangements. Results of these simulations show that dexterity is improved by using an independent revolute joint at the tip of a continuum robot instead of achieving distal rotation by transmission of rotation about the backbone of the continuum robot. Further, it is shown that designs with two robotic continuum robots as surgical arms have diminished dexterity if the bases of these arms are close to each other. This result justifies our design and points to ways of improving the performance of existing designs that use continuum robots as surgical arms.Keywords
This publication has 17 references indexed in Scilit:
- Design and Integration of a Telerobotic System for Minimally Invasive Surgery of the ThroatThe International Journal of Robotics Research, 2009
- Single-port laparoscopic surgery: an overviewSurgical Endoscopy, 2009
- Totally Laparoscopic Central HepatectomyJournal of Gastrointestinal Surgery, 2008
- An Investigation of the Intrinsic Force Sensing Capabilities of Continuum RobotsIEEE Transactions on Robotics, 2008
- Robotic system for no‐scar gastrointestinal surgeryThe International Journal of Medical Robotics and Computer Assisted Surgery, 2008
- Transvaginal laparoscopically assisted endoscopic cholecystectomy: a hybrid approach to natural orifice surgeryGastrointestinal Endoscopy, 2007
- Fisherman's Technique, Introducing a Novel Method for Using the Umbilical Port for Removal of Appendix During Laparoscopic AppendectomySurgical Laparoscopy, Endoscopy & Percutaneous Techniques, 2007
- Quantification of motion characteristics and forces applied to tissues during suturingThe American Journal of Surgery, 2005
- A micro optical force sensor for force feedback during minimally invasive robotic surgerySensors and Actuators A: Physical, 2004
- Medical robotics in computer-integrated surgeryIEEE Transactions on Robotics and Automation, 2003