Abstract
We present an instrumental realization of linearity measurement of optical detectors based on flux addition of two light-emitting diodes (LEDs) in an integrating sphere. The instrument measures the linearity ratios at pre-defined photocurrent levels according to the algorithm which eliminates the drift effect of LED flux (Shin D J et al 2005 Metrologia 42 154). As a result, the non-linearity correction factors for detector responsivity are determined with their relative standard uncertainties ranging from 2 × 10−5 to 6 × 10−5 (k = 1) for photocurrent ranges from 10−8 A to 10−3 A. By changing the type of LEDs, the wavelength dependence of non-linearity can also be measured, which is demonstrated at 390 nm, 670 nm and 750 nm for a Si photodiode used as a transfer standard for spectral responsivity scale comparisons.