The Effects of Loading on the Preload and Dimensions of the Abutment Screw for a 3-Unit Cantilever-Fixed Prosthesis Design
Authors:
- Hoda YousefHYHoda Yousef
- David EhrenbergDEDavid Ehrenberg
- Allyn LukeALAllyn Luke
- Saul WeinerSWSaul Weiner
- 1 August 2013
- journal article
- research article
- Published by Wolters Kluwer Health in Implant Dentistry
- Vol. 22 (4) , 414-421
Abstract
The purpose of this study was to use an in vitro model system to compare the effects on the screw torque and screw dimensions within 2 commercially available implant systems from occlusal loading on a cantilevered-fixed partial denture. Cantilevered implant-supported 3-unit prostheses with 2 premolar abutments and 1 premolar pontic (7.3 mm in length) were made on resin casts containing 2 implant analogs for 2 implant systems: BioLok Silhouette Tapered Implant System (Birmingham, AL) and Zimmer Tapered Screw-Vent Implant System (Carlsbad, CA) with 10 samples in each group. Each sample was loaded with either of 2 protocols: (1) a load of 50 N on the cantilevered pontic unit and (2) a loading of 150 N on all 3 units. The outcome measures were (1) changes in residual torque of the abutment screws and (2) changes in screw dimension. The BioLok Silhouette Tapered Implant group demonstrated slight but statistically significant torque loss 18.8% to 28.5% in both abutment screws for both protocols, P ≤ 0.05, without any changes in screw dimension. In the Zimmer Tapered Screw-Vent Implant group, there was a significant elongation of the abutment screws and a markedly significant 44.4%, (P ≤ 0.01) loss in torque in the mesial screw and a 28.5%, (P ≤ 0.05) loss in torque in the distal screw when the cantilever alone was loaded. Differences in screw design influence the maintenance of preload and distortion of the shank. The influence of the interface design, namely an internal hex of 1 mm versus an external hex did not influence the preload. Cantilevered prostheses can cause loss of torque and dimensional changes in abutment screws.Keywords
This publication has 30 references indexed in Scilit:
- Survival and complication rates of implant‐supported fixed partial dentures with cantilevers: a systematic reviewClinical Oral Implants Research, 2009
- Biomechanics/risk management (Working Group 2)Clinical Oral Implants Research, 2009
- Bone level changes at implants supporting crowns or fixed partial dentures with or without cantileversClinical Oral Implants Research, 2008
- The influence of implant location and length on stress distribution for three-unit implant-supported posterior cantilever fixed partial denturesThe Journal of Prosthetic Dentistry, 2004
- Implant‐supported fixed cantilever prostheses in partially edentulous arches. A seven‐year prospective studyClinical Oral Implants Research, 2003
- Implant-retained cantilever fixed prosthesis: Where and whenThe Journal of Prosthetic Dentistry, 2000
- Design of prosthetic cantilever bridgework supported by osseointegrated implants using the finite element methodDental Materials, 1998
- Finite element analysis of the effect of cantilever and implant length on stress distribution in an implant-supported fixed prosthesisThe Journal of Prosthetic Dentistry, 1996
- Cantilever and Implant Biomechanics: A Review of the Literature, Part 2Journal of Prosthodontics, 1994
- Effect of cantilever length on stress transfer by implant-supported prosthesesThe Journal of Prosthetic Dentistry, 1994