Increased loosening of cemented straight stem prostheses made from titanium alloys. An analysis and comparison with prostheses made of cobalt-chromium-nickel alloy
We studied the rate of aseptic loosening of three different types of femoral stems in primary total hip replacement. After a median follow-up of 10.2 years 4/147 CoCrNi (SS77) straight stems (type M.E. Müller) were revised. After a median follow-up of 7.7 years 32/239 Ti-6Al-7Nb (SS77) were revised and after a median follow-up of 5.2 years 52/203 SLS Titanium alloy stems were revised. Whereas the first two stems are of identical design (smooth-blasted, anterior and posterior collar), the SLS stem design is different. Surgical procedure and cementing technique have remained unchanged. There is a significantly higher risk of failure for smaller titanium stem sizes and in males and patients who are physically active. This indicates that the greater elasticity of the Titanium alloy is one of the factors responsible for loosening. Nous avons étudié le taux de descellement de trois différents types de tiges fémorales utilisées lors d’une arthroplastie primaire totale de la hanche. Après un suivi moyen de 10,2 années, 4 tiges droites de type CoCrNi (SS77) (type M.E. Müller) sur 147 ont été révisées. 32 tiges en alliage Ti-6A1-7Nb sur 239 ont été révisées après un suivi moyen de 7,7 années, alors que 52 tiges de type SLS Titane sur 203 ont été révisées après un suivi moyen de 5,2 années. Les deux premières tiges présentent le même type de design. Les tiges SLS sont quant à elles d’un autre type. La procédure chirurgicale et la technique de cimentation sont restées inchangées. Le risque d’échec est significativement plus grand pour les plus petites tiges en titane, chez les hommes ainsi que chez les patients physiquement actifs. Ceci montre que la plus grande élasticité de l’alliage en titane est un des facteurs responsable de descellement.