The accuracy of image-free computer navigation in the placement of the femoral component of the Birmingham Hip Resurfacing
Authors:
Open Access
- 1 April 2007
- journal article
- research article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 89 (4) , 557-560
Abstract
A cadaver study using six pairs of lower limbs was conducted to investigate the accuracy of computer navigation and standard instrumentation for the placement of the Birmingham Hip Resurfacing femoral component. The aim was to place all the femoral components with a stem-shaft angle of 135°. The mean stem-shaft angle obtained in the standard instrumentation group was 127.7° (120° to 132°), compared with 133.3° (131° to 139°) in the computer navigation group (p = 0.03). The scatter obtained with computer-assisted navigation was approximately half that found using the conventional jig. Computer navigation was more accurate and more consistent in its placement of the femoral component than standard instrumentation. We suggest that image-free computer-assisted navigation may have an application in aligning the femoral component during hip resurfacing.Keywords
This publication has 15 references indexed in Scilit:
- Computer Navigation versus Standard Instrumentation for TKAClinical Orthopaedics and Related Research, 2005
- Computer Assisted Navigation in Total Knee ArthroplastyThe Journal of Arthroplasty, 2005
- Randomized Control Trial Comparing Radiographic Total Knee Arthroplasty Implant Placement Using Computer Navigation Versus Conventional TechniqueThe Journal of Arthroplasty, 2005
- Computer-Assisted Navigation Increases Precision of Component Placement in Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2005
- Birmingham hip resurfacing arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2005
- Image-Based Computer-Assisted Total Knee Arthroplasty Leads to Lower Variability in Coronal AlignmentClinical Orthopaedics and Related Research, 2004
- ORIENTATION OF THE FEMORAL COMPONENT IN SURFACE ARTHROPLASTY OF THE HIPJournal of Bone and Joint Surgery, 2004
- Metal-on-metal resurfacing of the hip in patients under the age of 55 years with osteoarthritisThe Journal of Bone and Joint Surgery. British volume, 2004
- Metal-on-Metal Hybrid Surface Arthroplasty: Two to Six-Year Follow-up StudyJournal of Bone and Joint Surgery, 2004
- Reduction in Variability of Acetabular Cup Abduction Using Computer Assisted Surgery: A Prospective and Randomized StudyComputer Aided Surgery, 2002