Optically stimulated luminescencein vivodosimetry for radiotherapy: physical characterization and clinical measurements in60Co beams
Authors:
- I MrčelaIMI Mrčela
- T BokulićTBT Bokulić
- J IzewskaJIJ Izewska
- M BudanecMBM Budanec
- A FröbeAFA Fröbe
- Z KusićZKZ Kusić
- 26 August 2011
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (18) , 6065-6082
Abstract
A commercial optically stimulated luminescence (OSL) dosimetry system was investigated for in vivo dosimetry in radiation therapy. Dosimetric characteristics of InLight dot dosimeters and a microStar reader (Landauer Inc.) were tested in (60)Co beams. The reading uncertainty of a single dosimeter was 0.6%. The reproducibility of a set of dosimeters after a single irradiation was 1.6%, while in repeated irradiations of the same dosimeters it was found to be 3.5%. When OSL dosimeters were optically bleached between exposures, the reproducibility of repeated measurements improved to 1.0%. Dosimeters were calibrated for the entrance dose measurements and a full set of correction factors was determined. A pilot patient study that followed phantom validation testing included more than 100 measured fields with a mean relative difference of the measured entrance dose from the expected dose of 0.8% and the standard deviation of 2.5%. In conclusion, these results demonstrate that OSL dot dosimeters represent a valid alternative to already established in vivo dosimetry systems.Keywords
This publication has 38 references indexed in Scilit:
- Evaluation of a novel 4D in vivo dosimetry systemMedical Physics, 2009
- Temperature coefficients for in vivo RL and OSL dosimetry usingRadiation Measurements, 2008
- Optically stimulated luminescence and its use in medical dosimetryRadiation Measurements, 2006
- CW-OSL measurement protocols using optical fibre Al2O3:C dosemetersRadiation Protection Dosimetry, 2006
- Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancerRadiotherapy and Oncology, 2006
- The nonmonotonic dose dependence of optically stimulated luminescence in Al2O3:C: Analytical and numerical simulation resultsJournal of Applied Physics, 2006
- Real-time optical-fibre luminescence dosimetry for radiotherapy: physical characteristics and applications in photon beamsPhysics in Medicine & Biology, 2004
- Nonlinear dose dependence and dose-rate dependence of optically stimulated luminescence and thermoluminescenceRadiation Measurements, 2001
- Guidelines on the implementation of diode in vivo dosimetry programs for photon and electron external beam therapyMedical Dosimetry, 1999
- Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40Medical Physics, 1994