Monte Carlo calculations of , the beam quality conversion factor
Authors:
- B. R. MuirBMB. R. Muir
- D. W. O. RogersDRD. W. O. Rogers
- 27 October 2010
- journal article
- Published by Wiley in Medical Physics
- Vol. 37 (11) , 5939-5950
Abstract
Purpose: To use EGSnrc Monte Carlo simulations to directly calculate beam quality conversion factors,, for 32 cylindrical ionization chambers over a range of beam qualities and to quantify the effect of systematic uncertainties on Monte Carlo calculations of . These factors are required to use the TG‐51 or TRS‐398 clinical dosimetry protocols for calibrating external radiotherapy beams.Methods: Ionization chambers are modeled either from blueprints or manufacturers' user's manuals. The dose‐to‐air in the chamber is calculated using the EGSnrc user‐code egs_chamber using 11 different tabulated clinical photon spectra for the incident beams. The dose to a small volume of water is also calculated in the absence of the chamber at the midpoint of the chamber on its central axis. Using a simple equation, is calculated from these quantities under the assumption that is constant with energy and compared to TG‐51 protocol and measured values.Results: Polynomial fits to the Monte Carlo calculated factors as a function of beam quality expressed as and are given for each ionization chamber. Differences are explained between Monte Carlo calculated values and values from the TG‐51 protocol or calculated using the computer program used for TG‐51 calculations. Systematic uncertainties in calculated values are analyzed and amount to a maximum of one standard deviation uncertainty of 0.99% if one assumes that photon cross‐section uncertainties are uncorrelated and 0.63% if they are assumed correlated. The largest components of the uncertainty are the constancy of and the uncertainty in the cross‐section for photons in water.Conclusions: It is now possible to calculate directly using Monte Carlo simulations. Monte Carlo calculations for most ionization chambers give results which are comparable to TG‐51 values. Discrepancies can be explained using individual Monte Carlo calculations of various correction factors which are more accurate than previously used values. For small ionization chambers with central electrodes composed of high‐Z materials, the effect of the central electrode is much larger than that for the aluminum electrodes in Farmer chambers.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada
- Cummings Foundation
- Ontario Innovation Trust
This publication has 23 references indexed in Scilit:
- The determination of beam quality correction factors: Monte Carlo simulations and measurementsPhysics in Medicine & Biology, 2009
- The replacement correction factors for cylindrical chambers in high-energy photon beamsPhysics in Medicine & Biology, 2009
- Monte-Carlo-based perturbation and beam quality correction factors for thimble ionization chambers in high-energy photon beamsPhysics in Medicine & Biology, 2008
- Efficiency improvements for ion chamber calculations in high energy photon beamsMedical Physics, 2008
- Modelling ionization chamber response to nonstandard beam configurationsJournal of Physics: Conference Series, 2008
- Calorimetric determination ofkQfactors for NE 2561 and NE 2571 ionization chambers in 5 cm × 5 cm and 10 cm × 10 cm radiotherapy beams of 8 MV and 16 MV photonsPhysics in Medicine & Biology, 2007
- Wall correction factors, , for thimble ionization chambersMedical Physics, 2006
- Absorbed‐dose beam quality conversion factors for cylindrical chambers in high energy photon beamsMedical Physics, 2000
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999