Incidence of Ceramic Liner Malseating in Trident® Acetabular Shell
- 12 March 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 467 (6) , 1552-1556
Abstract
The low wear rates associated with ceramic hip articulations have made them a popular bearing for younger patients. Although few complications have been observed, one report revealed several instances of incomplete seating of the ceramic liner in the metallic shell. We performed a cohort study of consecutive THAs using a ceramic-ceramic bearing. Radiographic analysis showed 50 (7.2%) of the group of 694 hips had evidence of incomplete seating of the liner in the metallic shell. Although we observed no adverse effects at 6 to 12 weeks, we encourage surgeons to carefully assess liner placement in the metal shell at the time of surgery to avoid this unintended consequence and to assess placement at the time of followup so patients can be properly followed when incomplete seating is identified. Level of Evidence: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.Keywords
This publication has 14 references indexed in Scilit:
- Ceramic Liner Fracture after Cementless Alumina-on-Alumina Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2007
- Incomplete seating of the liner with the Trident acetabular systemThe Journal of Bone and Joint Surgery. British volume, 2007
- Acetabular Component Deformation with Press-Fit FixationThe Journal of Arthroplasty, 2006
- Ceramic Failure After Total Hip Arthroplasty with an Alumina-on-Alumina BearingJournal of Bone and Joint Surgery, 2006
- A Titanium-encased Alumina Ceramic Bearing for Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2005
- Concerns about Ceramics in THAClinical Orthopaedics and Related Research, 2004
- Failure of ceramic THR with liner dislocation – a case reportActa Orthopaedica, 2004
- Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cupThe Journal of Arthroplasty, 2003
- Ceramic-on-Ceramic Bearings in Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2002
- Ultra-low wear rates for rigid-on-rigid bearings in total hip replacementsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2000