Design and implementation of algorithms for focus automation in digital imaging time‐lapse microscopy
Authors:
- 26 November 2002
- Vol. 49 (4) , 159-169
Abstract
Background Digital time‐lapse microscopy using Nomarski‐DIC requires that an autofocusing system adapt to changes in cell shape, size, and position while contending with drift, noise, and hysteresis in the microscope and imager. We have designed and implemented an autofocusing system that tracks subjects under dynamic conditions and maintains focus within a threshold of discriminability. Methods With the use of proven and novel algorithms for autofocusing in Nomarski, we performed “virtual” experiments on recorded image stacks to simulate drift and sudden displacements and test the search algorithm response. Results We found that combining a simple [1, −1] contrast function with an adaptive “warmer–colder” focusing algorithm yields a reasonable compromise between focusing precision and noise tolerance. This method was implemented to record growth kinetics of yeast cells in single and multiple fields of view over several hours. Conclusions We have implemented a robust digital autofocus that maintains focus on optically complex samples imaged at high resolution. The tolerance of this system of drift and vibration suggests that it is a practical system for time‐lapse imaging in many biological applications. Cytometry 49:159–169, 2002.Keywords
This publication has 16 references indexed in Scilit:
- A two‐channel four‐dimensional image recording and viewing system with automatic drift correctionJournal of Microscopy, 2000
- Evaluation of autofocus functions in molecular cytogenetic analysisJournal of Microscopy, 1997
- Comparison of phase‐contrast and fluorescence digital autofocus for scanning microscopyCytometry, 1994
- Comparison of autofocus methods for automated microscopyCytometry, 1991
- FocusingInternational Journal of Computer Vision, 1988
- The RAD9 Gene Controls the Cell Cycle Response to DNA Damage in Saccharomyces cerevisiaeScience, 1988
- A comparison of different focus functions for use in autofocus algorithmsCytometry, 1985
- Computer-Oriented analysis of human chromosomes—III. FocusComputers in Biology and Medicine, 1972
- The Staircase-Method in PsychophysicsThe American Journal of Psychology, 1962
- Sequential minimax search for a maximumProceedings of the American Mathematical Society, 1953